The rear-panel manifolds can be easily changed while retaining their
connections. For example, if the Product is in a rack and must be removed for
service, the manifold can be taken out of the Product while all necessary hoses
and inputs are left connected.
The rear-panel manifold can be easily removed if the port threads become
damaged. To remove the manifold, loosen the four manifold bolts and pull the
manifold out. Replace the manifold and tighten the four bolts, torque: 6.2 N·m
(55 lbf in). See Figure 23.