Error Messages

MESSAGE
Appears when the batteries
E-b
are too low.
Appears when the
wrong code strip is
E-C
inserted or other coding
errors.
E-2
Expired code strip.
Appears when a used test
E-U
strip is inserted.
Appears when ambient
temperature is above or
E-t
below system operation
range.
E-0,
E-A,
Problem with the meter.
E-E.
Appears when test strip is
removed while counting
E-F
down, or insufficient blood
volume.
WHAT IT MEANS
WHAT TO DO
Replace the batteries
immediately.
Check if the code numbers on
the display and ketone foil
packet are the same. If the
problem persists, please
contact GlucoRx Customer care
on 01483 755133 for help.
Repeat the test with a new lot of
test strip.
Repeat with a new test strip.
System operation range is
10°C to 40°C (50°F to 104°F).
Repeat the test after the meter
and test strip are in the above
temperature range.
Repeat the test with a new test
strip. If the meter still does not
work, please contact GlucoRx
Customer care on 01483
755133 for assistance.
Review the instructions and
repeat test with a new strip.
If the problem persists, please
contact GlucoRx Customer care
on 01483 755133 for help.