Error Indicators - ReliOn BP300 Instruction Manual

Hide thumbs Also See for BP300:
Table of Contents

Advertisement

SYMBOL
Air plug disconnected.
Arm cuff not applied
correctly.
Air is leaking from the
arm cuff.
Cuff over or under
inflated.
The arm cuff was inflated
above 299 mmHg when
inflating the cuff manually.
Movement during
measurement.
Clothing is interfering
with the arm cuff.
Device error.
Batteries are worn.

ERROR INDICATORS

CAUSE
CORRECTION
Insert the plug securely. Refer to
"Applying the Arm Cuff".
Apply the arm cuff correctly. Refer to
"Applying the Arm Cuff".
Replace the cuff with a new one.
Repeat measurement. Remain still and do
not talk during measurement. Refer to
"Taking a Measurement".
If "E2" appears repeatedly, inflate the cuff
manually until it is 30 to 40 mmHg above
your previous measurement result. Refer
to "Taking a Measurement".
Do not inflate the cuff above 299 mmHg.
Refer to "Taking a Measurement".
Repeat measurement. Remain still and do
not talk during measurement. Refer to
"Taking a Measurement".
Remove any clothing interfering with the
arm cuff. Refer to "Applying the Arm
Cuff".
Contact Customer Service.
Replace the four batteries.
Refer to "Battery Installation".