Figure 88 Selecting loop suppression verification
3

Setup considerations for stable RF oscillator measurement

Measurement noise floor
The signal amplitude at the test set's R input (Signal Input) port sets the
measurement noise floor level. Use
page 135 to determine the amplitude required to provide a noise floor level
that is below the expected noise floor of your DUT. (The Checking the Beatnote
procedure in this section will provide you with an opportunity to estimate the
measurement noise floor that your DUT will provide.)
134
e5505a_user_select_loop
24 Jun 04 rev 3