WIRELESS TIG FOOT PEDAL

diagnostiC mode

The system's range is maximized with a clear line of sight to the
power source. Other factors such as heavy RF activity within the
pedal's vicinity can also affect performance.
The diagnostic mode will provide a qualitative analysis of the
system's communication. In this mode, the battery strength
indicator will change to illuminate BLUE, and blink at a rate of 20
times per second to indicate uninterrupted communication. An
intermittent blinking rate would suggest a heavily congested
environment, or that the pedal and receiver are approaching their
maximum operating distance. (See figure A.10)
FIGURE A.10
BATTERY STRENGTH
BLUE LED INDICATOR

diP switCh FunCtions

Each pedal and receiver has a 2 position dip switch and a 16
position rotary dip switch. The functionality of each is as follows:
Mode of Operation
Control
Diagnostic
Configure
Calibration
X-NA
TWO GREEN
LEDS

PRINTED CIRCUIT BOARD

Pedal
Receiver
Dip Switches
Dip Switches
#1
#2
#1
off
off
off
off
off
off
on
x
on
x
on
x
Entering Diagnostic Mode:
(See figure A.11)
1. Remove the 4 screws on the back of the receiver and remove
the cover.
2. Set DIP SWITCH #2 on the receiver board to "ON."
3. Replace cover.
FIGURE A.11
#2
off
Normal Transmission
on
Communication strength metric
x
x
Set/Clear dip at maximum stroke
B-5
OPERATION
BACK COVER REMOVED
AND SCREWS FOR CLARITY.
SET DIP
SWITCH #2
TO ON
1 2
ON
OFF
Details/Function