7.2 Standard Specifications

INSTRUMENT RANGE
Hematocrit (Hct): 10 Hct – 60 Hct
Oxygen Saturation: 55% – 100%
(Hct 18)
OPERATING AND STORAGE
TEMPERATURE
50º F to 104º F
(10º C to 40º C)
Avoid extreme temperatures during
transportation (<32º, >110 Fº)
HEMATOCRIT ACCURACY
10 Hct – 60 Hct: ± 1 Hct SD
OXYGEN SATURATION ACCURACY
55% – 100%: ± 2 Sat % Hct 18
ACCESS BLOOD FLOW
Estimates flow rates
50ml/min – 2500ml/min ± 15%
RECIRCULATION CAPABILITY
Estimates recirculation values>4%
BATTERY CAPACITY
2 hours continuous on full charge
FULL CHARGE TIME
36 hours
PHYSICAL DIMENSIONS
5.25" H, 8.25" W, 11.63" L, 5 lbs.
HD BLOOD FLOW RATE
CAPABILITY
50 ml/min – 1300 ml/min
INTERNAL DATA STORAGE
Sufficient to store up to 26 hours of
data indefinitely
Performance Characteristics
7-4
INPUT
12 VDC/1 Ampere/12 watts
1.2A max
POWER SUPPLY UNIT
Universal regulated supply
100VAC – 240VAC
12 VDC/1.25 Ampere Output
COMMUNICATIONS PORTS
8-position keyed modular jack
(Keyed RJ-45) serial port
DB female parallel port
ELECTRICAL SHOCK PROTECTION
Class II
Internally powered
Type BF
WATER PROTECTION
Splash proof
ANESTHETIC SUITABILITY
Not suitable
MODE OF OPERATION
Continuous
BATTERY
Type: Rechargeable sealed lead
battery (12V, 2A-h) Battery is replaceable
Polarity: ± (labeled on battery)
Mode of insertion: Via 2 pin
Molex connector (replaceable)