Section 4 Ending A Session

This section explains how to end a monitoring session. Also included in
this section are instructions on how to use the output options on the CLM
III. The headings in this section are as follows:
____________________________________________

4.1 Ending the Dialysis Session

PROFILE
STARTUP
Monitoring with the CLM III should be concluded prior to the rinse-back procedure,
stopping the blood pump, or moving the sensor clip back to the verification filter.
Otherwise, the CLM III may indicate a false Hct reading. If the CLM III does read a
false Hct, it will cause the blood volume and Hct graphs to rescale. Rescaling makes
the information on the graph difficult to interpret. Properly ending a monitoring
session will prevent unnecessary rescaling.
To end a monitoring session:
Ending a Session
Ending the Dialysis Session......................................... 4.1
Output Options........................................................ 4.2
SETUP
OUTPUT
Press MENU or SELECT
4-1