SPORT ELEC MULTISPORTPRO User Manual page 9

4 channels
Hide thumbs Also See for MULTISPORTPRO:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 7
III. USING ThE UNIT
STEP 1
Take the electrodes and remove carefully their protection, dampen them one by one with water on your fingertips on the adhesive
gel, and wait for 2 minutes (diagram I. A, page 2).
STEP 2
Fit the electrodes on your body as shown in diagram II. A or B or C, page 2 : Positions are shown for information only
Important: For optimal use, carefully read each page and look at the examples of use (photos II, page 2).
Examples of use (photos II, page 2)
= connector channel B (middle blue) ;
= connector channel C (light blue) ; SUP = upper limbs ; INF = lower limbs
B
photo - Chest
1 - Large chest
2 - Medium deltoid
Do not place the electrodes
3 - Posterior
close to the heart
4 - Anterior
STEP 3
Take connector (CEV8) (photo I. B, page 2) and connect it :
1 - to the electrodes (photo I. C, page 2)
2 - then to the unit (photo I. D, page 2).
STEP 4
The unit is connected to the electrodes through the connector, composed of 2 groups of 4 x 1m. long wires completed by 2 groups
of 4 connection plugs.
It is possible to hang your MULTISPORT PRO to your clothing using the clip.
STEP 5
Start your unit following the "Starting the unit" procedure described in paragraph II.
Contractions should be firm but never painful. If you feel an unpleasant sensation around the electrodes : bring the power-on key
back to minimum, move the electrode(s) and gradually increase the power.
Once the adjustments have been optimised in this way, leave your appliance to work for the full duration of the program.
STEP 6
Once the program is complete, disconnect the electrodes from the (CEV8) connector, then the latter from the unit.
Then pull off the electrodes gently, dampen them slightly with a few drops of water, then return them carefully onto their support.
Store the adhesive electrodes in a fresh location, in a plastic bag (e.g. vegetable container in the refrigerator).
IV. RECOMMENDaTIONS OF USE
The unit is recommended for physiotherapists for functional re-education.
• You can restart a program's cycle on another muscle zone or change the program while remaining on the same muscle zone,
without exceeding however 40 min use per muscle zone and per day.
• For optimum use, refer to the diagrams shown in this leaflet.
• Clean the unit and its cables with a cloth dampened with a disinfecting solution.
• The adhesive electrodes can be rehydrated before and after use with a disinfecting agent (e.g. Hexamedine).
• The adhesive electrodes must be repositioned on their support, and stored in a fresh storage between +5°C and +10°C.
• The adhesive electrodes have a lifetime of about 40 uses, you can order them directly from the online shop www.sport-elec.com,
or from your reseller.
14
C
photo - Shoulders
photo - Trapezius
9 - Upper trapezius
10 - Medium trapezius
PRECAUTIONS OF USE
Do not use the unit on your face
f
f
Do not leave within reach of children
f
f
Ask for advice from your physician:
if you have electronic or metallic implants (pins...)
f
f
in case of skin condition (wounds...)
f
f
wARNINGS
Keep away from children or people with mental disabilities.
f
f
The device contains small pieces that can be swallowed.
In order to prevent any risk of strangulation, please keep
f
f
the electrodes wires away from children.
If the device is not likely to be used for some time, the
f
f
batteries should be removed
Multisportpro is meant for personnal use. For reasons of
f
f
hygiene, the electrode should only be used on one person.
Do not use the device by driving or by working on a
f
f
machine.
Do not use this electro stimulation unit in a damp
f
f
environment, a bathroom, a sauna, or close to electric
machinery connected to mains and earthed, or even
piping earthed.
Use this device in a clean place (without dust nor dirt..).
f
f
The simultaneous connection of a patient to a high
f
f
frequency surgical unit may cause burns at the points of
contact of the stimulator's electrodes, and the stimulator
may be damaged.
Operation in close proximity of a short wave therapy unit
f
f
may cause instability in the stimulator's output power.
Do not, under any circumstances, attempt to open or
f
f
repair the Multisportpro yourself! Only have repairs carried
out by the customer service department.
CUSTOMER SUPPORT: +33 ( 0 ) 2 32 96 50 50
V. TEChNICaL SPECIFICaTIONS
ELECTRODES
Electrodes
Adhesive, hydrophilic and hypoallergenic
. . . . . . . . . . . . .
Size
89 x 50 mm
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Estimated lifetime
40 uses
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Class I medical device in accordance with Directive 93/42/CEE
UNIT:
Class IIa medical device in accordance with Directive 93/42/CEE
Software version
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Type of current
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Nomber of programs
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Number of program variations (or phases)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Number of adjustable independent outputs
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maximum intensity on a 500 Ω load/channel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TROUbLE SHOOTING
If your device doesn't work properly, do not use it and
contact your retailer.
RECOMMENDATIONS
Please apply regulations enforced in your country
when disposing your product at its end of life.
Therefore, please discard it in a location planned to
this effect, ensuring it is disposed in an environment-
friendly manner.
CONTRA-INDICATIONS
Do not use SPORT-ELEC®:
On the heart area
f
f
If you carry a pacemaker
f
f
If you are pregnant
f
f
During first 6 weeks after baby birth
f
f
In case of neuromuscular diseases
f
f
In case of hemorrhagic diseases
f
f
On the face and skull
f
f
CARE AND CLEANING INSTRUCTIONS
Do not expose to bleach.
f
f
Clean the unit with a tissue dampened with a disinfecting
f
f
solution diluted.
Keep the electrodes in a dry and cool place. After use,
f
f
moisten the adhesive gel surface with a few drops of
water and hold two minutes.
ELECTRODES wARRANTY CONDITIONS
The electrodes are for a single person use only. For hygiene
reasons, they cannot be returned nor replaced.
E
Dual phase
8 "SPORT" programs - 6 "HEALTH" programs
344
4
100 mA / 500 Ω / channel
15

Advertisement

Table of Contents
loading

Table of Contents