Problem
Motor will not turn
Motor will not turn –
overcurrent condition
The pump will not stay on the
motor
Troubleshooting Guide
Solution
1. Make sure the tubing is loaded
properly. The tubing should be
centered in the middle of the rollers.
Reload if necessary.
2. Check to see if the motor is turned
on.
3. Check motor fuse.
1. Make sure the tubing is loaded
properly. The tubing should be
centered in the middle of the rollers.
2. Make sure the tubing wall thickness
is correct (refer to Table 9-1).
3. The tubing durometer may be too
high for the pump (refer to Chapter
9).
4. The tubing may be caught on the
bottom roller. Reposition the roller
slightly and load the tubing again.
1. The pump is not properly mounted
to the motor. Refer to Section 3.
2. The mounting screws are loose.
Tighten the screws, if necessary.