waRnings :
• If the device is not likely to be used for some time, the
batteries should be removed
• Do not use the device when driving or working on a
machine.
• Do not use this electro stimulation unit in a damp environ-
ment, a bathroom, a sauna, or close to electric machinery
connected to mains and earthed, or even piping earthed.
• The simultaneous connection of a patient to a high
frequency surgical unit may cause burns at the points of
contact of the stimulator's electrodes, and the stimulator
may be damaged. Operation in close proximity of a short
wave therapy unit may cause instability in the stimulator's
output power.
RecoMMendaTions :
Please apply regulations enforced in your
country when disposing your product at its end of life. There-
fore, please discard it in a location planned to this effect,
ensuring it is disposed in an environment-friendly manner.
v. TEchnIcAL spEcIFIcATIOns
unit :
Type of current ......................................................................................................................... Dual phase
Number of programmes ..................................................................................................... 8 programs
Number of program variations (or phases) .............................................................. 55
Number of adjustable independent outputs ......................................................... 2
Current .......................................................................................................................................... From 0 to 70 mA - From 0 to 1000 ΩΩ
Frequency range / Pulse width range ......................................................................... 1 to 100 Hz / 160 to 320 µs
Power ............................................................................................................................................. 3 x 1,5 V alkaline batteries type LR6 (not provided)
Unit dimensions and weight ............................................................................................ 80 x 160 x 35 mm - 200 gr
Automatic stop at end of each program ................................................................... Yes - about 4 min 30 sec
Activation safety ...................................................................................................................... 100%
Absence of electrodes or of contact detection ..................................................... yes detection independent on each of the 2 outputs
Power indicator ........................................................................................................................ bar luminous graph
Weak and/or defective batteries indicator ............................................................... yes - luminous indicator
Absence contact electrodes Indicator ........................................................................ yes - luminous indicator
Indicator of the programs .................................................................................................. luminous pictograms
Temperature of storage / Relative humidity ........................................................... 0°C to + 45°C / Between 10% and 90%
Operating temperature / Relative humidity ............................................................ + 5°C to + 45°C / 20% to 65%
Connection device / electrodes ..................................................................................... by wires (about 90 cm) - plugs ø 2 mm
Caution .........................................................................................................................................
RoHS ................................................................................................................................................ The unit is free of materials likely to have a negative impact
Guarantee ................................................................................................................................... 2 years (unit only)
BeLT :
Size ......................................................................................................Single size
(about 150 cm long) / (i.e. 2 half-belts 75cm)
Weight ...........................................................................................................150g
Fabric ........................................................polyester laminated on foam
Built-in electrodes....................................................carbonated silicon.
Estimated lifetime .....................................................around 12 months
guaRanTee :
The belt, cream, and electrodes are designed for personal use.
For reasons of hygiene, there will be no trade-in nor exchange.
16
BeLT MAiNTAiNiNg AND CLeANiNg reCOM-
MendaTions.
- No washing
- No solvent
- No drying
- No ironing
- No laundering
- Clean the belt using a cloth dampened with a detergent,
then wipe with a dry cloth.
- Clean the unit with a tissue dampened with a disinfecting
solution (Mercryl type) diluted with water at 1/10th.
- Clean the electrodes with a mix of water/hypoallergenic
detergent, then rinse thoroughly.
RadiaTed elecTRic Field
Electric field at 3m.
Frequency (Mhz)
(dB(μV/m)
200,0
< 30
400,0
< 30
600,0
< 30
800,0
< 30
1000,0
< 30
Read the manuel thoroughly
on the environment
sTandaRds
Belt ...................................................................................................... ISO 10993
Contact cream , electro-conducting ................75 ml / 2.54 Fl Oz
Electrodes........................................................................................ ISO 10993
Unit ................................................................EN60601-1 / EN60601-2-10
Classification ........................................................................................type BF
BODY BEAUTIFUL
BEschrEIBUng DEs gEräTEs
4
10
2
5
6
vOr DEr BEnUTZUng DEs gEräTEs LEsEn
sie haben ein leistungsfähiges und starkes profigerät erworben.
1 - Bitte lesen Sie die Bedienungsanleitung aufmerksam durch.
2 - Vor Gebrauch stellen Sie Batterien in Ihrem Gerät.
3 - Testen Sie das Gerät an den Bizeps gemäß dem nachfolgend beschriebenen Schnelltest.
FUnKTIOns-schnELLTEsT
1 - Schließen Sie 2 Haftelektroden an Ihr Gerät an (sieh Verbindungsverfahren in der Gebrauchsanweisung).
2 - Testen Sie Ihr Gerät am Programm PG02.
3 - Steigern Sie allmählich die Leistung.
4 - Sie beginnen, Zusammenziehungen zu spüren, das Gerät funktioniert.
5 - Nach dieser Entdeckungstour wissen Sie, wie Sie die Leistungen des Gerätes für sich nutzen können..
cUsTOmEr sUppOrT : 00 33 2 32 96 50 50
3
7
10
1
1
Taste ON/OFF
2
Programmwähltaste.
3
Anzeige des laufenden Programms.
4
Anzeige der Leistung : kanal A.
8
5
Leistung steigern von kanal A (Ebene 0 bis 30).
6
Leistung senken von kanal A (Ebene 30 bis 0).
7
9
Anzeige der Leistung : kanal B.
8
Leistung steigern von kanal B (Ebene 0 bis 30).
9
Leistung senken von kanal B (Ebene 30 bis 0).
10
Abschaltung : blinkt auf, wenn die Elektroden
abgeschaltet werden.
17