Chapter 3
LEDs, DIP Switches, and Connectors

SAFE MODE Switch

NI CVS-1450 Series User Manual
Figure 3-3 shows the location of the DIP switches on the CVS-1450 device.
ON
To start the CVS-1450 device in Safe mode, move the SAFE MODE
switch to the ON position and reset the CVS-1450 device. Use Safe mode
to reconfigure TCP/IP settings and to download or update software from
the development computer.
Downloading incorrect software to the CVS-1450 device may cause it to
hang during restart or become inaccessible over the network. Powering on
or resetting the CVS-1450 device in Safe mode starts the CVS-1450 device
but does not start the embedded LabVIEW RT engine. To resume normal
operations, restart the CVS-1450 device with the SAFE MODE switch in
the OFF position.
Figure 3-3. DIP Switches
3-4
NI 1454
Compact Vision System
ni.com