TEMPERATURE RECORDER (OPTION)
5. Fix the recorder sensor to the inside of the sensor
cover (MTR-C8) by using the enclosed clips and
screws and seal the access port on the chamber side
with a silicon. (Fig. 3)
6. Seal the access port on the outside with putty as
shown in Fig. 4.
7. Replace the front panel and fix it with 4 screws.
8. Operate the freezer until the chamber temperature
gets to the set temperature.
temperature and chamber temperature displayed on
the control panel. Adjust the zero adjustment volume
on the temperature recorder so that the recorded
temperature can corresponds with the displayed
temperature if they are not compliance each other.
Refer to the instruction manual enclosed with the
recorder.
Check the recorded
34
Recorder sensor
Sensor cover
Putty
Silicon
Fig. 3
Fig. 4