Patient (Source) Leakage Current Test - GE B30 Technical Reference Manual

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor

2.9 Patient (source) leakage current test

This procedure only applies to Class I (grounded/earthed) equipment, and measures the
leakage current from the ECG/RESP connector or the SPO2 connector of the device to ground.
DMM
NOTE: The MD represents the network and voltage measuring instrument and its frequency
characteristics per IEC 60601-1.
The patient connector test body shorts all signals in the connector together. Refer to the
instructions contained with the safety analyzer to perform this test.
(1)
Connect the ECG/RESP Test Body to the green connector of the device under test.
(2)
Configure leakage tester as follows:
(3)
Power on Device under test.
(4)
Read and record the current leakage indicated on the tester.
(5)
Change leakage tester switches to:
(6)
Read and record the current leakage indicated on the tester.
(7)
Change leakage tester switches to:
(8)
Read and record the current leakage indicated on the tester.
(9)
Change leakage tester switches to:
3-10
.
Document no
2044677-001
Leakage Tester
HIGH
LOW
GND
Closed
0.15µF
1K
10
DMM set to measure AC voltage
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
GND (Earth) - CLOSED
Polarity - NORMAL
Neutral - OPEN
NORM
Power Cord
RVS
GND
ECG Test Body
or SPO2 Test Body
Device
Under
Test