GE B30 Technical Reference Manual page 56

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
Table 4
SpO
2
Table 5
Invasive blood pressure
connectors (Dual BP)
1-42
.
Document no
2044677-001
SpO
connector
2
connector
Invasive blood pressure connectors (P1, P2)
Pin No.
Signal
1
DET_A
2
DET_C
3
DATA-
4
Wire 1/3
5
IR_C
6
OUTER SHIELD
7
DET_SHIELD
8
PRB_ID
9
Wire 3/5
10
RED_C
11
DATA+
Pin
Signal
Description
No.
1
BP_+V
BP transducer excitation voltage, channel
REF
1
2
BP SIG+
BP transducer signal positive (+), channel
1
3
BP_+V
BP transducer excitation voltage, channel
REF
2
4
AGND
Analog ground
5
BP SIG+
BP transducer signal positive (+), channel
2
6
SHIELD
BP cable shield
7
AGND
Analog ground
8
BP SIG1
BP transducer signal negative (-), channel
1
9
BP SIG2
BP transducer signal negative (-), channel
2
10
BP1_ID
BP1 probe identification
11
NC
Not connected
Description
Photodiode anode
Photodiode cathode
LED connection
IR LED cathode
Bin/ID Resistor+
LED Connection
RED LED cathode

Hide quick links:

Advertisement

Table of Contents
loading