GE B30 Technical Reference Manual page 216

Patient monitor
Hide thumbs Also See for B30:
Table of Contents

Advertisement

B30 Patient Monitor
Recorder
17. Paper compartment cleaning
Open the paper compartment hatch and take out the paper roll, if installed.
Remove any paper chaff from the compartment.
Clean the thermal printhead and the small glass window in front of the static brush with a
cotton swab dipped in isopropyl alcohol. Avoid contact with the rubber paper roller.
NOTE: Be careful to limit the application of alcohol to the thermal printhead and the window.
Leave the paper compartment empty and close the hatch.
Configure the monitor screen so that all required parameters are shown, for example:
Plug in the Hemodynamic Module. Connect a patient simulator to the module and check
that all connected parameters are shown on the screen.
Preset recording settings:
Record/Print
Monitor Setup
Default Trend - Graph.
Display Trends - Time Scale - 2 h
18. Paper recognition
Press the
Out of paper' appears on the screen.
19. Cover state recognition
Open the paper compartment cover. Check that the previous message changes to
'Recorder: Cover open'.
Install a paper roll and close the cover. Check that the message 'Recorder: Cover open'
disappears from the screen.
8-16
.
Document no
2044677-001
Monitor Setup
- Screen 1 Setup - Waveform Fields - Field 1 - ECG1
-
Record Waveforms - Waveform 1 - ECG1
Record Trends - Graphic Trend 1 - Hr
Graphic Trend 2 - P1
-
Install/Service (password 16-4-34) - Installation - Printer & Recorder -
Start/Stop
key on the monitor side panel. Check that the message "Recorder:
Field 2 - ECG2
Field 3 - P1
Field 4 - P2
Field 5 - PLETH
Field 6 - OFF
Waveform 2 - P1
Waveform 3 - P2
Delay - Off
Paper Speed - 6.25 Mm/S
Length --> 30 S