Ultrasonic flaw detector
Table of Contents

Advertisement

6 Documentation
You will then automatically switch to the first operating
level with the A-scan.
Next to the A-scan, the control functions are displayed
in the function group RECORD, whereas the settings for
each gate in question are displayed in the function
groups of gates.
Recording information is displayed on top of the A-scan:
● TIME = current time of recording,
● FRAM = number of single frames.
6-18
– In the function group RECORD, select the function
RECORD and press one of the two selector keys to
start the recording.
– Switch to the function FLAG and press one of the two
selector keys to set a flag. The name of the flag cor-
responds to the number of the single frame when the
function key is pressed.
– Switch to one of the function groups of gates and op-
erate the gates as usual.
– Vary the gain as usual.
– In the function group RECORD, select the function
RECORD and press one of the two selector keys to
interrupt the recording.
– In the function group RECORD, select the function
STOP and press one of the two selector keys. The re-
cording is stopped, and you will see the normal dis-
play view of the first operating level with the A-scan
and the function groups.
Issue 2 (12/2013)