5 Operation
– Select the function gate a start and position the gate
A on the first DAC echo.
– Select the function AUTO80 and press one of the se-
lector keys to automatically set the echo to 80 %
screen height.
– Select the function RECORD and press one of the
selector keys to store the first DAC point. The record-
ing is confirmed by the indication (1 POINT). After
this, the function gate a start is automatically select-
ed.
5-106
Distance-amplitude curve according to JIS Z3060-2002 (JISDAC)
– Position the gate A on the second DAC echo.
– Select the function AUTO80 and press one of the se-
lector keys to automatically set the echo to 80 %
screen height.
– Select the function RECORD and press one of the
selector keys to store the second DAC point. The re-
cording is confirmed by the indication (2 POINTS). Af-
ter this, the function gate a start is automatically
selected again.
– Record further DAC points in the same way.