nIBp C
alIBratIon
This function allows the calibration of the internal pressure
measurement system.
Select „NIBP Calibration" in the service menu and confirm
by pressing NIBP.
Offset calibration
During this calibration NO pressure should be present in
the system.
The cuff connector (self closing) has to be opened (use cuff
connector).
1. Remove cuff.
2. Connect the cuff connector with a short tube.
3. The actual pressure value is displayed.
4. Select SET and confirm by pressing NIBP.
5. The zero value is averaged and stored
Calibration with 200 mmHg
Connect the complete measurement system (hand pump,
manometer and compensating container) as described in
„NIBP test".
6. Using the hand pump, apply a fixed pressure of
200 mmHg to the system
(control by the external manometer).
7. Wait several seconds until the pressure is stable;
correct the pressure if necessary.
8. Select SET and confirm by pressing NIBP.
9. The NIBP calibration is finished
Select EXIT (and confirm by pressing NIBP) to return to the
service menu.
2018111-336 Rev. B
NIBP Calibration
Offset:
SET
EXIT
p
reSSure calibration
NIBP Calibration
Apply
SET
EXIT
p
reSSure calibration
eBike III basic / comfort
0
mmHg
-
offSet
200
mmHg
-
gain
- 77 -