Dynamic Load Calibration - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

a
/ C
lIgnMent
alIBratIon
d
l
C
ynaMIC
oad
alIBratIon
Dismount BOTH (!) pedal cranks of the eBike III - see
"Dismount pedal cranks" under section „Dismount
pedal cranks" on page 41.
Follow the operator manual from the manufacture
of calibration device being used to connect the
calibration device with the axis of the eBike III.
When a dynamic calibration is performed, the measured
actual load values are entered into the ergometer software.
The ergometer calculates the exact characteristic load
curve for regulating the eddy current brake.
The following load / rpm combinations have to be
measured during the calibration process:
offset
at
50 Watt
at
50 Watt
at
400 Watt
at
To activate the service menu the ergometer has to be
switched off and on again.
As soon as the self test appears on the ergometer display,
press the two softkeys synchronously:
To prevent settings from being changed by chance, the
service menu is „protected" by a password.
Select „999" with the arrow keys and confirm with NIBP.
Navigate through the menus by using the arrow keys
(saddle height) to position the cursor bar.
Confirm the selection by pressing NIBP (to the right
of the LCD).
- 72 -
no load / no rotations
50 rotations / min
110 rotations / min
110 rotations / min
eBike III basic / comfort
Note
This session describes the dynamic load calibration of the
eBike III as a manual „step-by-step" procedure
When using the ergoline calibration bench ("ergoline
ergotest 550" - order direct from Ergoline P/N 705890), an
automatic dynamic load calibration is possible -
please refer to the operator manual of the ergotest 550
device
Service Code
999
p
aSSword for Service menu
2018111-336 Rev. B