Appendix D. Modbus Register Map
Function
Parameter
Channel 5 /
Output A
Channel 5 /
Output B
Channel 6 /
Output A
128
Table 13: Modbus Register Map (cont.)
Sub Parameter
ID
ModBus
Address
Type
426
Zero
428
Span
430
Output (percent
500
of scale)
State
502
Units
504
Type
506
Zero
508
Span
510
Output (percent
520
of scale)
State
522
Units
524
Type
526
Zero
528
Span
530
Output (percent
600
of scale)
State
602
Type
Access Modbus
2426 32-bit Integer
RW
2428 32-bit float
RW
2430 32-bit float
RW
2512 32-bit float
RW
2502 32-bit Integer
RW
2504 32-bit Integer
RW
2506 32-bit Integer
RW
2508 32-bit float
RW
2510 32-bit float
RW
2532 32-bit float
RW
2522 32-bit Integer
RW
2524 32-bit Integer
RW
2526 32-bit Integer
RW
2528 32-bit float
RW
2530 32-bit float
RW
2612 32-bit float
RW
2602 32-bit Integer
RW
Qty. of
Notes
Request
Registers
Read Holding
2
1 = 4-20 mA,
(0x03)
2 = 0-20 mA,
3 = 0-2 volts
Read Holding
2
(0x03)
Read Holding
2
(0x03)
Read Holding
2
When TEST MODE
(0x03)
enabled, force output
to value
Read Holding
2
2 = TEST MODE,
(0x03)
All else = run mode
Read Holding
2
Measurement/Unit
(0x03)
Code 0 = off,
others see tab
Read Holding
2
1 = 4-20 mA,
(0x03)
2 = 0-20 mA,
3 = 0-2 volts
Read Holding
2
(0x03)
Read Holding
2
(0x03)
Read Holding
2
When TEST MODE
(0x03)
enabled, force output
to value
Read Holding
2
2 = TEST MODE,
(0x03)
All else = run mode
Read Holding
2
Measurement/Unit
(0x03)
Code 0 = off,
others see tab
Read Holding
2
1 = 4-20 mA,
(0x03)
2 = 0-20 mA,
3 = 0-2 volts
Read Holding
2
(0x03)
Read Holding
2
(0x03)
Read Holding
2
When TEST MODE
(0x03)
enabled, force output
to value
Read Holding
2
2 = TEST MODE,
(0x03)
All else = run mode
moisture.IQ User's Guide