Chapter 9. Specifications
9.9.1c
Process Temperature
-110°C to +70°C (-166°F to +158°F) dew/frost point temperature
9.9.1d
Storage Temperature
70°C (158°F) maximum
9.9.1e
Operating Pressure
5 microns of Hg to 5000 psig
9.9.1f
Flow Range
Gases:
Static to 10,000 cm/sec linear velocity at 1 atm
Liquids:
Static to 10 cm/sec linear velocity at 1 g/cc
Response Time: <5 seconds for 63% step change in moisture content in either wet up or dry down cycle.
9.9.1g
Moisture Image Series Probe/Analyzer Separation
915 m (3000 ft) with provided cable.
9.9.1h
Moisture Image Series Probe/Analyzer Cable
Unshielded, twisted pair, maximum loop resistance of 100 ohms
9.9.1i
Built-in Temperature Sensor (optional)
Type:
Thermistor network
Operating Range: -30°C to +70°C (-22°F to +158°F)
Accuracy:
±0.5°C overall
9.9.1j
Built-in Pressure Sensor (optional)
Type:
Solid state/piezoresistive
Available Ranges: 30 to 300 psig
50 to 500 psig
100 to 1000 psig
300 to 3000 psig
500 to 5000 psig
Accuracy:
±1% of span
102