Setting Up Input And Output (I/O); Connecting The Analyzer To A Network Via Ethernet; Configuring The Analyzer's Ip Address - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 6: Analyzer Operation

Setting Up Input and Output (I/O)

Connecting the Analyzer to a Network Via Ethernet

The Analyzer can be connected to your network or directly to a personal computer via an Ethernet cable. If you
connect the Analyzer to your network, use a standard Ethernet cable. If you connect the Analyzer directly to a
personal computer, use an Ethernet crossover cable.
N
:
Consult with your Information Technology group to determine the best configuration for your
OTE
networking environment.

Configuring the Analyzer's IP Address

You must configure the Analyzer's Internet Protocol (IP) settings for use in connecting the Analyzer to a
computer network or directly to a computer.
To configure the Analyzer's IP Address
1.
Select the I/O tab, and then press the IP Configuration button. The Configuration screen appears.
2.
Press one of the following buttons to configure the Analyzer's IP address.
• Off (default) — Indicates that the Analyzer is not communicating with a network or other computer.
• Automatic — The Analyzer is connected to your network with a standard ethernet cable, and your
network uses a DHCP server to dynamically assign IP addresses.
GE Analytical Instruments ©2013
Figure 37: IP Configuration Screen
137 of 238
DLM 68100-06 EN Rev. A