Please specify the following parameters when ordering a transmitter; process type, process
range (%, ppm), thermocouple type, temperature scale F/C, analog output 1 process and
scale, analog output 2 process and scale.
Calibration
Function
Cold Junction
Thermocouple
min
Thermocouple
max
Millivolt
Millivolt
Analog 1 Zero
Analog 1 Span
Analog 2 Zero
Analog 2 Span
Calibration
Function
Cold Junction
Thermocouple
Min
Thermocouple
Max
Millivolt
Millivolt
Analog 1 Zero
Analog 1 Span
Analog 2 Zero
Analog 2 Span
11/14/2006
NOTE:
Typical Oxygen Transmitter Calibration
(F840030)
Measured Value or
Input
Room Temp
800°F (B type)
standard t/c type
3000°F (B type)
standard t/c type
0.0 mV
2000 mV
0% O2
20.9% O2
800°F +/- 5°
3000°F +/- 5°
Typical Carbon Transmitter Calibration
(F840031)
Measured Value or
Input
Room Temp
MUST BE
SPECIFIED
MUST BE
SPECIFIED
0.0 mV
2000 mV
0% Carbon
2.55% Carbon
MUST BE
SPECIFIED
MUST BE
SPECIFIED
Page 1 of 23
Output / Units
°F
°F
°F
Millivolts
Millivolts
4.0 mA +/- 0.1
20.0 mA +/- 0.1
4.0 mA +/- 0.1
20.0 mA +/- 0.1
Output / Units
°F
°F
°F
Millivolts
Millivolts
4.0 mA +/- 0.1
20.0 mA +/- 0.1
4.0 mA +/- 0.1
20.0 mA +/- 0.1
Rev. 14