HEX
DEC
PARAMETER
14
20
SPARE
15
21
SPARE
16
22
SPARE
17
23
TEMPFIL
HEX
DEC
PARAMETER
18
24
MVFIL
19
25
AZERO
1A
26
ANUM
1B
27
BZERO
1C
14
BNUM
1D
15
PROC
1E
16
COLDJCT
1F
17
TEMP
20
18
MV
21
19
DACV1
11/14/2006
BLOCK 0
DESCRIPTION
source value is based on the selection in
ASRC upper byte where
SPARE
SPARE
SPARE
Temperature Input Filter in seconds
Range = 0 to 3276. The higher the number
the faster the reading update.
DEFAULT = 1000
BLOCK 1
DESCRIPTION
Millivolt Input Filter in seconds
Range = 0 to 3276. The higher the number
the faster the reading update.
DEFAULT = 1000
LINEAR OFFSET, Y INTERCEPT LINEAR
SCALING FOR INPUT A
LINEAR SPAN VALUE FOR INPUT A
LINEAR OFFSET, Y INTERCEPT LINEAR
SCALING FOR INPUT B
LINEAR SPAN VALUE FOR INPUT B
This value is the calculated process value
shown as an integer. The decimal point and
exponent values are required to determine the
actual scaled value.
Range = -999 to 9999.
For example: If the process = oxygen, display
decimal point = 2, and exponent = 6, and
PROC = 1234, then the actual value and
displayed as 12.34 ppm.
COLD JUNCTION
Where 1 COUNT = 1°F (°C), RANGE = -99 TO
255°F (°C). Note this parameter is an
unsigned integer.
MEASURED TEMPERATURE
Where temperature is presented in degrees C
or F, based on the C/F setting. Note this
parameter is an unsigned integer of
temperature -2721 = 62815
Range = max / min range of selected
thermocouple.
MEASURED MILLIVOLT
Where this value is scaled in 0.1 mV
increments, i.e. 10001 = 1000.1.
Range = 0 to 2000 mV.
ANALOG OUTPUT 1
0 to 4095 is 4 to 20 mA In dual mode 4mA = -
100, 12mA = 0, 20mA = +100
Page 16 of 23
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ/WRITE
READ ONLY
READ ONLY
READ ONLY
READ ONLY
READ/WRITE
Rev. 14