Card Configuration; Preamplifier Module - Keithley 1801 Instruction Manual

Keithley 1801 nanovolt preamp
Table of Contents

Advertisement

Installation
2.2
Card configuration
2.2.1 Preamplifier module
Figure 2-1 shows the configuration of the preamplifier
module. Key items include:
INPUT Terminals
The HI and LO INPUT terminals are pure copper studs
and nuts intended for connecting input signals to the
Model 1801. To minimize thermal EMFs, use only pure
copper lugs or wires for connections, and be sure that both
the terminals and connecting lugs are clean and free of
oxidation. See paragraph 2.4.4 for details on input
connections.
CAUTION
The
maximum
INPUTS HI and LO is 1V @ 100mA
peak (inputs over 2mV require one-
minute recovery). The maximum volt-
1
Figure 2-1
Preamplifier module configuration
2-2
signal
between
KEITHLEY
1801 NANOVOLT PREAMP
INPUTS
HI
2mV
PEAK
LO
41V
PEAK
Input
Terminals
age between LO and chassis ground is
41V peak. Exceeding these limits may
result in damage.
Offset Adjustments
V ZERO and I ZERO are externally accessible
adjustments for nulling voltage and current offsets
respectively. These controls need not be adjusted during
normal operation.
Improper
degrade performance. The offset adjust-
ments should only be performed using
the procedures covered in Section 5.
Power Supply Connector
This connector attaches the preamplifier module to the
power supply card using the supplied cable. See para-
graph 2.4.1 for details.
V ZERO
I ZERO
CAUTION:
!
NO INTERNAL OPERATOR SERVICEABLE PARTS,
SERVICE BY QUALIFIED PERSONNEL ONLY.
2
Zero
Adjustments
NOTE
offset
adjustments
will
Power Supply
3
Connector

Advertisement