Pince multimètre
Clamp multimeter
Vielfachmesszange
Pinza multimetro
Pinza multimétrica
F R A N Ç A I S
E N G L I S H
D E U T S C H
I T A L I A N O
E S P A N O L
Notice de fonctionnement
User's manual
Bedienungsanleitung
Libretto d'Istruzioni
Manual de Instrucciones
F 05

Advertisement

Table of Contents
loading

Summary of Contents for Chauvin Arnox F 05

  • Page 1 F 05 Pince multimètre Clamp multimeter Vielfachmesszange Pinza multimetro Pinza multimétrica F R A N Ç A I S Notice de fonctionnement E N G L I S H User's manual D E U T S C H Bedienungsanleitung I T A L I A N O Libretto d’Istruzioni...
  • Page 2 English Meaning of symbol Caution! Please consult the User Manual before using the device. In this User Manual, failure to follow or carry out instructions preceded by this symbol may result in personal injury or damage to the device and the installations. Meaning of symbol This appliance is protected by double insulation or reinforced insulation.
  • Page 3: Table Of Contents

    CONTENTS PRESENTATION ..............27 DESCRIPTION ............... 27 IMPLEMENTATION - FUNCTIONAL CHARACTERISTICS ... 30 Reference conditions ............30 Voltage measurements (V) ..........31 Audio continuity test ( ) ..........32 Resistance measurement (Ω) ......... 32 Semi-Conductor Test ( ) ..........33 Current measurements (A) .........
  • Page 4: Presentation

    1. PRESENTATION Clamp multimeter F05 emphasises reliability and simplicity of use to respond to the needs of power professionals: A compact unit integrating the current sensor for intensity measurements without breaking the test circuit Outstanding ergonomic features, in particular: - automatic or manual selection of the type of signal to be measured, AC or DC, - automatic selection of measurement calibre, - programmable audio voltage indication: V-Live,...
  • Page 5 Command keys The keys are capable of 3 types of action: Short pressure < 1.3 s, it is valid if the key pressure is detected. Long pressure > 1.3 s, this gives access to a measurement or operating mode. Holding or releasing the key has no effect. Held pressure Gives access to a measurement or operating mode and remains in this mode as long as pressure is held.
  • Page 6 HOLD key/ switch combination (see description § 3.12): - Suppression of automatic stop function - Activation of the V-Live function - Display of the internal software version Yellow key/ switch combination (see description § 3.12): - Modification of audio indication threshold in continuity test - Default configuration of the unit MIN/MAX key/ switch combination (see description §...
  • Page 7: Implementation - Functional Characteristics

    PEAK ON in V and A in MIN/MAX mode if the measurement of the peak value is selected Indicates the display of a maximum value in MIN/MAX mode Indicates the display of a minimum value in MIN/ MAX mode Symbol which goes on during the phase rotation direction detection sequence.
  • Page 8: Voltage Measurements (V)

    - Peak factor of applied AC signal: √ 2 - Position of conductor in clamp jaws: centred. - Conductor diameter: ≤ 5 mm - No external AC magnetic field - No electrical field 3.2 Voltage measurements (V) 1. Connect the measurement leads to the instrument’s terminals, complying with the polarities indicated: red lead on the “+”...
  • Page 9: Audio Continuity Test ( )

    3.3 Audio continuity test ( 1. Connect the measurement cables to the unit terminals. 2. Set the rotary switch to position " ". 3. Connect the unit to the circuit to be tested. The buzzer is continuously active as soon as contact is established (circuit closed) and if the resistance value measured is lower than the threshold value chosen by the programming (adjustable from 1 to 40 Ω, see §...
  • Page 10: Semi-Conductor Test ( )

    Display range 400 Ω 4000 Ω 40 kΩ Measuring range 0.0 to 400 to 4.00 kΩ to 399.9 Ω 3999 Ω 39.99 kΩ Accuracy (1) : 1% R + 2 pt Resolution 0.1 Ω 1 Ω 10 Ω Open circuit voltage ≤...
  • Page 11: Inrush Function

    Range switching and AC/DC selection are automatic. Actuate the yellow key to force AC/DC selection manually if necessary. Correction of the current measurement zero (DC zero) To measure the low currents, perform a zero correction first. - Press and hold the HOLD key until zero appears on the display.
  • Page 12: Power Measurement (W)

    The field of application is limited to industrial frequencies (15 Hz... 70 Hz) Implementation This function is accessible in AC current measurement only, after selection of the MIN/MAX mode. Action Display Comments Press 0.5 P then the value the yellow for rms Enter the function corresponding...
  • Page 13: Power Factor Calculation (Pf)

    Active power measurement characteristics Display range 4000 W 40 kW 400 kW Measuring range (2) 5 to 4.00 kW to 40.0 kW to 3999 W 39.99 kW 240.0 kW (1) Accuracy (3) 2% R + 1 pt Resolution 10 W 100 W 1) the scale is limited to 240 kW in one-phase (600 V x 400 A).
  • Page 14: Phase Order Indication

    1. Perform a short pressing on the Hz key, the display shows the frequency of the measured signal. 2. Pressing again returns to the previously displayed measurement. Characteristics Display range 40 Hz 400 Hz 4000 Hz 40 kHz Measuring range (1) 10.00 to 40.0 to 400 to...
  • Page 15: Secondary Functions

    Note 3: The measurement principle is based on a certain frequency stability and practically sinusoidal signals (THD < 10%). This excludes in particular measurement on power generators whose spin stabilisation system is too weak to ensure adequate frequency stability. Action Display Comments Switch on...
  • Page 16: Automatic Compensation For Cable Resistance

    3.12.3 Automatic compensation for cable resistance This compensation is obtained by holding down the HOLD key when the continuity test or measurement resistance function is selected. When the key is released, when the display shows zero, the correction value is memorised in the clamp. If the value measured is higher than 2 Ω...
  • Page 17: Activation Of The V-Live Function

    3.12.9 Activation of the V-Live function Hold the HOLD key pressed and bring the rotary switch from the OFF position to the V position. The unit emits a double beep then the V and symbol flashes. The selected configuration is memorised when the key is released: the V symbol is then ON steady and the symbol flashes.
  • Page 18: General Specifications

    4. GENERAL SPECIFICATIONS 4.1 Dimensions and weight 70 x 193 x 37 mm 260 g 4.2 Clamp Tightening Capacity ≤ 26 mm 4.3 Power supply a standard 9 V alkaline battery (type CEI 6LF22, 6LR61 or NEDA 1604) Average battery charge life: 60 h or 20,000 x 10 s measurements Battery charge indicator Flashing: charge life <...
  • Page 19: Variations In Operating Range

    Electric shocks (test as per IEC 1000-4-5) - 6 kV in differential mode on the voltmeter function, aptitude criterion B - 2 kV induced on the current measurement cable , aptitude criterion B Electromagnetic compatibility (as per EN 61326-1 ed. 97 + A1) Emission: class B Immunity: - Electrostatic discharges:...
  • Page 20 Quantity Meas. range Quantity Influence quantities quantities influenced Typical Maxi Adjacent Conductor conductor in contact with crossed by external 45 dB 40 dB a current perimeter 400 A jaws Conductor clamped by 0...400 A DC < 1 ct 1 ct the clamp or RMS Application...
  • Page 21: Marginal Operating Conditions

    Typical frequency response curve - V f (f) 8.00% 6.00% 4.00% 2.00% 0.00% -2.00% Specified limits -4.00% -6.00% -8.00% -10.00% 10000 Hz 100 Hz 10 Hz 1000 Hz Frequency (Hz) - I f (f) 10.00% 5.00% 0.00% Specified limits -5.00% -10.00% -15.00% -20.00%...
  • Page 22: To Order

    5. TO ORDER F05 ..............P01.1209.05Z Delivered in a blister with set of 2 probe cables, 1 cable with male safety connector, 1 alligator clip, 1 x 9 V battery, one carrying case and this operating manual Accessories and spare parts set of 2 cables with probe tips (NF EN 61010) ......
  • Page 23: Maintenance

    7. MAINTENANCE For maintenance, use only specified spare parts. The manufacturer will not be held responsible for any accident occuring following a repair done other than by its After Sales Service or approved repairers. 7.1 Changing the battery The clamp must be disconnected from any external electrical source and must not be in contact with the cable.