Chauvin Arnox C.A 6165 Quick Start Manual

Appliance multitester

Advertisement

Quick Links

GB -
Quick start guide
C.A 6165
Appliance multitester

Advertisement

Table of Contents
loading

Summary of Contents for Chauvin Arnox C.A 6165

  • Page 1 GB - Quick start guide C.A 6165 Appliance multitester...
  • Page 2: Table Of Contents

    Contents General description ............................3 Warnings and notes ........................... 3 1.1.1 Safety warnings ............................. 3 1.1.2 Warnings related to safety of measurement functions ................3 1.1.3 Markings on the instrument ........................4 Instrument description ..........................5 Front panel ..............................5 Instrument operation ............................
  • Page 3: General Description

    1.1.1 Safety warnings In order to reach high level of operator safety while carrying out various measurements using the C.A 6165 instrument, as well as to keep the test equipment undamaged, it is necessary to consider the following general warnings: ...
  • Page 4: Markings On The Instrument

     DO NOT touch any part of test probe in front of the barrier (keep your fingers behind the finger guards on the probe) – possible danger of electric shock!  It is a good practice to use lowest possible trip-out current. 1.1.2.2 Diff.
  • Page 5: Instrument Description

    2 Instrument description 2.1 Front panel C.A 6165 Figure 2.1: Front panel Mains supply connector F1, F2 fuses (F 5 A / 250 V) F3, F4 fuses (T 16 A / 250 V) On / Off switch Test connections TC1 for external test adapters...
  • Page 6 USB connector MicroSD card slot...
  • Page 7: Instrument Operation

    3 Instrument operation The C.A 6165 can be manipulated via a keypad or touch screen. 3.1 General meaning of keys Cursor keys are used to: select appropriate option Enter key is used to: confirm selected option start and stop measurements...
  • Page 8: Symbols And Messages

    3.3 Symbols and messages Supply voltage warning Possible causes:  No earth connection.  Instrument is connected to an IT earthing system. Press YES to continue normally or NO to continue in a limited mode (measurements are disabled). Warning: The instrument must be earthed properly to work safely! Resistance L-N >...
  • Page 9 In pre-test a too high external voltage was detected between ISO/SUB and PE terminals. The measurement was cancelled. Press OK to continue. In pre-test a possible high leakage current was detected. It is likely that a dangerous leakage current (higher than 3.5 mA) will flow after applying power to the device under test.
  • Page 10 Red dot indicates phase of measurement where higher leakage was measured. Applicable only if phase reversal is enabled during the measurement. Test leads resistance in Continuity P/S - PE measurement is not compensated. Test leads resistance in Continuity P/S - PE measurement is compensated.
  • Page 11: Single Tests

    4 Single tests 4.1 Single test measurements 4.1.1 Visual inspections Figure 4.1: Visual inspection menu Test circuit Figure 4.2: Visual inspection test circuit Visual inspection procedure  Select the appropriate Visual inspection.  Start the inspection.  Perform the visual inspection of the appliance / equipment. ...
  • Page 12: Continuity

    Figure 4.3: Examples of Visual inspection results 4.1.2 Continuity Figure 4.4: Continuity test menu Test results / sub-results R ....Resistance ΔU………..Voltage drop scaled to 10 A Test parameters Output connections Output [4-wire, P-PE] Test current I out [0.2 A, 4 A, 10 A, 25 A] Duration Duration [Off, 2 s ...
  • Page 13 Test circuit Figure 4.5: Measurement of continuity 4-wire Figure 4.6: Measurement of Continuity P/S - PE Continuity measurement procedure  Select the Continuity function.  Set test parameters / limits.  Connect test leads to C1, P1, P2 and C2 terminals on the instrument (4 wire), or connect test lead to P/S terminal (2 wire measurement P/S –...
  • Page 14 4.1.2.1 Compensation of test leads resistance This chapter describes how to compensate the test leads resistance in Continuity (Output = P/S – PE) function. Compensation can be carried out to eliminate the influence of test leads resistance and the internal resistances of the instrument on the measured resistance.
  • Page 15: Hv Ac

    4.1.3 HV AC IMPORTANT SAFETY NOTE Refer to chapter 1.1 Warnings and notes for more information regarding safe use of the instrument. Figure 4.10: HV AC test menu Test results / sub-results I ....test current U ....measured a.c. test voltage Ir ....
  • Page 16: Hv Dc

    HV AC measurement procedure  Select the HV AC function.  Set test parameters / limits.  Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.  Connect HV test leads to device under test.  Start measurement. ...
  • Page 17 Test limits High limit (I) H limit [0.05 mA ... 10.0 mA ] Low limit (I) L limit [Off, 0.05 mA ... 10.0 mA] Test circuit Figure 4.14: HV DC measurement HV DC measurement procedure  Select the HV DC function. ...
  • Page 18: Hv Ac Programmable

    4.1.5 HV AC programmable IMPORTANT SAFETY NOTE Refer to chapter 1.1 Warnings and notes for more information regarding safe use of the instrument. In the HV AC programmable test the time dependency of high voltage can be set according to diagram on Figure 4.16. Figure 4.16: Voltage / time diagram of the HV AC programmable test Figure 4.17: HV AC programmable test menu Test results / sub-results...
  • Page 19: Hv Dc Programmable

    Test circuit Figure 4.18: HV AC programmable test HV AC programmable test procedure  Select the HV AC programmable function.  Set test parameters / limits.  Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.  Connect HV test leads to device under test.
  • Page 20 Figure 4.20: HV DC programmable test menu Test results / sub-results U ....measured test voltage I ....test current Ic ....capacitive portion of test current Ir ....resistive portion of test current Test parameters Starting DC test voltage U start [500 V ...
  • Page 21: Insulation Resistance (Riso, Riso-S)

    HV DC programmable test procedure  Select the HV DC programmable function.  Set test parameters / limits.  Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.  Connect HV test leads to device under test. ...
  • Page 22 Test circuits Figure 4.24: Measurement of insulation resistance (ISO(+), ISO(-)) Figure 4.25: Measurement of insulation resistance (Socket LN - PE) Figure 4.26: Measurement of Riso, Riso-S (socket)
  • Page 23: Sub-Leakage (Isub, Isub-S)

    RISO measurement procedure  Select the Riso function.  Set test parameters / limits.  Connect test leads to ISO(+), ISO(-) terminals on the instrument, then connect test leads to device under test,  Connect device to mains test socket. For Riso-S test, additionally connect test lead to P/S terminal on instrument, and then connect test lead to device.
  • Page 24 L Limit (Isub) L limit [Off, 0.25 mA ... 15.0 mA, Custom ] H Limit (Isub-S) H limit [Off, 0.25 mA ... 15.0 mA ] L Limit (Isub-S) L limit [Off, 0.25 mA ... 15.0 mA ] Test circuits Figure 4.29: Measurement of Sub-leakage (SUB1, SUB2) Figure 4.30: Measurement of Sub-leakage (socket LN-PE) Figure 4.31: Measurement of Sub-leakage, Sub-leakage-S (socket) Sub-leakage measurement procedure...
  • Page 25: Differential Leakage

     Connect device under test to mains test socket. For Isub-S test, additionally connect test lead to P/S terminal on the instrument, and then connect test lead to a device.  Start measurement.  Measurement can be stopped manually or by timer. ...
  • Page 26 Test limits H Limit (Idiff) H limit [Off, 0.25 mA ... 15.0 mA, Custom ] L Limit (Idiff) L limit [Off, 0.25 mA ... 15.0 mA, Custom ] Output connections [Socket L,N – PE,P/S ]...
  • Page 27: Ipe Leakage

    Test circuit Figure 4.34: Measurement of Differential Leakage current Differential Leakage measurement procedure  Select the Differential Leakage function.  Set test parameters / limits.  Connect device under test to mains test socket and optionally to P/S terminal.  Start measurement.
  • Page 28 Test results / sub-results Ipe ....PE current P ....Power Test parameters Duration Duration [Off, 2 s ... 180 s] Change status Change [YES, NO] YES: The instrument measures leakage current in two sequential steps with delay* in between. The phase voltage is firstly applied to the right live output of the mains test socket and secondly to the left live output of the mains test socket.
  • Page 29: Touch Leakage

    Figure 4.38: Examples of Ipe Leakage measurement results 4.1.11 Touch Leakage Figure 4.39: Touch Leakage test menu Test results / sub-results Itou ....Touch Leakage current P ....Power Test parameters Duration Duration [Off, 2 s ... 180 s] Change status Change [YES, NO] YES: The instrument measures leakage current in two sequential steps with delay* in between.
  • Page 30: Power

    Test circuit Figure 4.40: Measurement of Touch Leakage current Touch Leakage measurement procedure  Select the Touch Leakage function.  Set test parameters / limits.  Connect device under test to mains test socket. Connect test lead to P/S terminal on the instrument and on device under test.
  • Page 31 Test results / sub-results P ....Active power S ....Apparent power Q ....Reactive power PF ....Power factor THDu .... Total harmonic distortion – voltage THDi ..... Total harmonic distortion – current Cos Φ ... cosinus Φ I ....Load current U ....
  • Page 32: Leak's & Power

    4.1.13 Leak's & Power Figure 4.45: Leak’s & Power measurement menu Test results / sub-results P ....Active power Itou ....Touch Leakage current Idiff ....Differential Leakage current S ....Apparent power Q ....Reactive power PF ....Power factor THDu ....
  • Page 33 Test circuit Figure 4.46: Measurement of Leak’s and Power Leak’s & Power measurement procedure  Select the Leak’s & Power function.  Set test parameters / limits.  Connect device under test to mains test socket and optionally to P/S terminal. ...
  • Page 34: Discharging Time

    4.1.14 Discharging Time Figure 4.48: Discharging Time test menu Test results / sub-results t ....Discharging time Up ....Peak voltage of supply during the test Test parameters Limit voltage Limit U [34 V, 60 V, 120 V] Output connections Output [External, Socket] Test mode Mode [Manual, Auto]...
  • Page 35 (1) peak voltage (4) Ulim (2) voltage at disconnection time (5) moment of disconnection (3) calculated voltage value (6) discharging time Figure 4.49: Measuring principle (external) Test circuit (Output = External) Figure 4.50: Discharging Time test (Output = External) Discharging Time test procedure (Output = External) ...
  • Page 36 Figure 4.51: Examples of Discharging Time measurement results (Output = External) Measuring principle (Output = Socket) The measuring principle of the Discharging time function is as following:  Phase The DEVICE UNDER TEST is connected to the mains test socket.The instrument monitors the mains voltage and internally stores the peak voltage value.
  • Page 37: Functional Inspections

    Figure 4.53: Examples of Discharging Time measurement results (Output = Socket) 4.1.15 Functional inspections Figure 4.54: Functional inspection start menu (left) and menu during inspection (right) Test parameters (optional) For the optional Power measurement test the parameters and limits are the same as set in the Power single test, see chapter 4.1.12 Power.
  • Page 38 Figure 4.56 Examples of Functional Inspection results...
  • Page 39: Maintenance

    5 Maintenance 5.1 Fuses There are four fuses on the front panel: F1, F2: F 5 A / 250 V / (20 × 5) mm / 1500 A: intended for instrument protection. For position of fuses refer to chapter 2.1 Front panel. F3, F4: T 16 A / 250 V / (32 ×...

Table of Contents