Cirrus CR:514 User Manual

Acoustic calibrator
Hide thumbs Also See for CR:514:

Advertisement

Quick Links

CR:514 & CR:515 Acoustic Calibrator
Page 1
CR:514 & CR:515
Acoustic Calibrator
User Manual
This manual, the software to which it relates, the
program code and drawings are all:
© Copyright Cirrus Research plc 1989-2013

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CR:514 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Cirrus CR:514

  • Page 1 CR:514 & CR:515 Acoustic Calibrator Page 1 CR:514 & CR:515 Acoustic Calibrator User Manual This manual, the software to which it relates, the program code and drawings are all: © Copyright Cirrus Research plc 1989-2013...
  • Page 2 The content of this manual, any illustrations, technical information and descriptions within this document were correct at the time of going to print. Cirrus Research plc reserves the right to make any changes necessary, without notice, in line with the policy of continuing product development and improvement.
  • Page 3: Table Of Contents

    Appendix 2 – Free Field Correction ................9 Microphone Correction Values ....................9 Example ..........................9 Appendix 3 - CE Declaration of Conformity ..............10 Appendix 4 Type Approval Certification ..............11 Warranty Information....................13 Cirrus Research Offices ............Error! Bookmark not defined.
  • Page 4: Operation

    Page 4 CR:514 & CR:515 Acoustic Calibrator Operation. Switching on the Calibrator Press the Power Button on the end of the Calibrator to switch the unit on. The Indicator will illuminate to show that the unit is operating. The calibrator will automatically switch off after 5 minutes to preserve battery power.
  • Page 5: Calibrating A Sound Level Meter

    CR:514 & CR:515 Acoustic Calibrator Page 5 Calibrating a Sound Level Meter. Push the microphone of the Sound Level Meter into the cavity at the end of the calibrator. Ensure the microphone is fully inserted into the cavity and is past the ‘O’ ring seals.
  • Page 6: Changing The Battery

    CR:514 & CR:515 Acoustic Calibrator Changing the Battery The CR:514 & CR:515 acoustic calibrators use a single 9v alkaline battery. This type of battery is known as 6F22 or NEDA 1604. It is also commonly known as PP3. Unscrew the screw holding the battery cover on, using a coin.
  • Page 7: Specification

    Page 7 Specification. Frequency 1kHz ± 1% Sound Level 94dB re 20μPa Standardisation CR:514 - IEC 60942:2003 Class 2 CR:515 - IEC 60942:2003 Class 1 Distortion Less than 2% Operating Humidity 25 to 90% Relative Humidity Operating Static Pressure 65 kPa to 108kPa...
  • Page 8: Appendix 1 - Technical Information

    Page 8 CR:514 & CR:515 Acoustic Calibrator Appendix 1 – Technical Information The normal mode of operation of the calibrator is with the unit switched on. When the LED indicates the unit is switched on this produces the greatest radio frequency emissions.
  • Page 9: Appendix 2 - Free Field Correction

    The correction is typically -0.3dB for ½ inch microphones (making the effective calibration level 93.7dB). The table below shows the correction values for the standard microphones of Cirrus Research plc. Calibration corrections are listed below for the Cirrus Research plc ½" Capsules and...
  • Page 10: Appendix 3 - Ce Declaration Of Conformity

    Page 10 CR:514 & CR:515 Acoustic Calibrator Appendix 3 - CE Declaration of Conformity Cirrus Research plc Hunmanby UK CE Certificate of Conformity Manufacturer: Cirrus Research plc Acoustic House, Bridlington Road Hunmanby, North Yorkshire, YO14 0PH United Kingdom Telephone +44 1723 891655...
  • Page 11: Appendix 4 Type Approval Certification

    CR:514 & CR:515 Acoustic Calibrator Page 11 Appendix 4 Type Approval Certification...
  • Page 12 Page 12 CR:514 & CR:515 Acoustic Calibrator...
  • Page 13: Warranty Information

    12. In the event of any dispute on the terms of the warranty Cirrus Research plc will accept pendulum arbitration by the United Kingdom Institute of Acoustics Ltd.
  • Page 14: Cirrus Research Offices

    CR:514 & CR:515 Acoustic Calibrator Cirrus Research Offices The addresses given below are the Cirrus Research plc offices. Cirrus Research plc also have approved distributors and agents is many countries worldwide. For details of your local representative, please contact Cirrus Research plc at the address below. Contact details for Cirrus Research authorised distributors and agents are also available from the Internet Web site at the address shown below.