Archos Blood Pressure Monitor User Manual page 29

Hide thumbs Also See for Blood Pressure Monitor:
Table of Contents

Advertisement

Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
2.333
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the device is used exceeds the applicable RF compliance
level above, the device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than[V1] V/m.
Not
applicable
3V/m
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic
site survey, a should be less than the compliance level
in each frequency range. b Interference may occur in
the vicinity of equipment marked with the following
symbol:
EN
29