TROUBLESHOOTING THE
The blood glucose level is higher than
525 mg/dL.
The blood glucose level is lower than
20 mg/dL.
Error message that indicates that there
is a problem with the meter, e.g.,
measurement error (time out, overflow,
offset) or temperature out of range.
Error message that indicates that there
may be a problem with the test strip, e.g,
the test strip may be damaged, moved,
or removed during testing, or inserted
improperly.
Error message could be caused by a used
or damaged test strip.
Error message indicates serial
communications error.
The battery sign appears on the display
with the unit of measurement. The power
of the battery is getting low. You can
complete about 50 more tests from the
time this symbol first appears.
Chapter 9
®
Xtra
This message indicates very high
blood sugar. You should recheck the
blood glucose level. If "Hi" again,
follow facility protocol.
This message indicates very low
blood sugar. You should recheck the
blood glucose level. If "Lo" again,
follow facility protocol.
Review the instructions and try again
with a new test strip. If the problem
persists, contact Customer Service.
Check the test strip for damage and
retest as necessary. Repeat the test.
If the error message appears again,
contact Customer Service.
Repeat the test with a new strip.
If the error message appears again,
contact Customer Service.
If error persists, contact Customer
Service.
Test results will be accurate but
replace the battery as soon as
possible. Battery type is
CR2032 3 volt.