Chapter 3
CONTROL SOLUTION TESTING
5. Compare the results with the ranges of
expected results shown on the test strip
vial. (Low = Blue Cap, High = Red Cap)
6. You should obtain results within the
expected range printed on the test
strip vial. If this is not so, repeat the test.
If the results are out of the range printed
on the test strip vial, check the
following:
Was the vial at room temperature?
Did you shake the bottle of control
solution before using?
Has the control solution expiration or discard date expired?
Is the meter malfunctioning?
7. Repeat steps 1-6 for the High
Control Solution procedure.
10
Cambridge Sensors USA, LLC
If the test result is still out of range,
call your Microdot® Customer
Service Representative at:
Toll Free (877) 374-4062
DO NOT test blood until you obtain
control results within the expected
ranges.