Error indications on the display.
(See Guide Book p. 37-41 for more information)
Indication
Cause
Test TIP is inserted at an angle (not
straight), or Test TIP is not fully inserted.
Used Test TIP is attached on the Meter.
Measuring window is dirty or dusty.
Please also refer to p. 50-53 "Meter operation check"
Measuring window may be dirty or dusty.
Clean the measuring window with a
cotton swab.
Please also refer to p. 50-53 "Meter operation check"
Environmental temperature is out of
specifi cation (10-35°C).
The measuring environment is too bright.
Point the meter at a darker place or move
to a place not exposed to direct sunlight.
" " indication will disappear.
Blood glucose value is higher than
33.3 mmol/L.
Blood glucose value is lower than
1.1 mmol/L.
Error indications on the display.
(See Guide Book p. 37-41 for more information)
Indication
Cause
Test TIP came off during measurement.
The measuring environment is too bright.
Test TIP came off during measurement.
The Meter is out of order. Contact Terumo.
Indication blinks
The batteries are almost empty.
Batteries are empty.
Measurement cannot be executed.
Indication disappears
2 minutes passed without any action
(In case the Test TIP is not on the Meter.)
5 minutes passed without any action
(In case the Test TIP is on the Meter.)
No indication on the display
The
"
"
button was not pressed.
The battery is placed incorrectly.
The batteries are empty.
mmol/L
Blood Glucose Monitoring System
for Self-Testing
SHORT INSTRUCTIONS FOR USE
Carefully read detailed instructions in the guide book
before using the blood glucose meter for the fi rst
time.
STEP 1 – PREPARATION
Wash and dry your hands thoroughly prior to use

Advertisement

loading
Need help?

Need help?

Do you have a question about the FineTouch and is the answer not in the manual?

Questions and answers

Summary of Contents for Terumo FineTouch

  • Page 1 fi rst Environmental temperature is out of time. Tel: The Meter is out of order. Contact Terumo. specifi cation (10-35°C). In case of emergency, please contact: STEP 1 – PREPARATION The measuring environment is too bright.
  • Page 2 STEP 2 – ATTACH TEST TIP STEP 3 – BLOOD SAMPLING STEP 4 – MEASURING STEP 5 – CLEAN UP Remove the Press the “ ” button Turn the dial to Make sure the display shows “ ” Remove the Test TIP and turn off the Meter Shallow Deep protective cap...

This manual is also suitable for:

Finetouch mmol/l