-------------------------------------------------------------------- CCP-3100 Opreator’s Manual All rights are reserved. Under copyright laws, this manual may not be copied, in whole or in part, without the prior written consent of HUVITZ Co., Ltd. Document Number: 9000OMCP001-A, published in 2006/07/20 CONTENTS Introduction ......................5 1.1.
Page 5
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- Input and Use of Program ..................18 7.1. How to input user's program ................18 7.2. How to use your stored programs..............19 Types of Chart (Type C is Standard)..............20 Maintenance and Repair..................21 9.1. Lamp Replacement..................21 9.2. Fuse Replacement...................22 The battery of Remote Controller Replacement …..……………………….23...
1.1. Outline of product HUVITZ CCP-3100 is an automatic chart projector which provides 41 different charts including red/green and polarized filters for the eye optometry. Users can select the desired charts instantaneously by cordless remote controller. Charts move quickly and quietly.
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 2.Safety Information 2.1 Introduction Safety is everyone’s responsibility. The safe use of this equipment is largely dependent upon the installer, user, operator, and maintainer. It is imperative that personnel study and become familiar with this entire manual before attempting to install, use, clean, service or adjust this equipment and any associated accessories.
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual 2.2 Safety Symbols The International Electrotechnical Commission (IEC) has established a set of symbols for medical electronic equipment, which classify a connection or warn of any potential hazards. The classifications and symbols are shown below. Save these instructions I and O on the power switch represent ON and OFF respectively.
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 2.3 Environmental factors Avoid the following environments for operation or storage: Where the equipment is exposed to water vapor. Don’t operate an equipment with a wet hand. Where the equipment is exposed to direct sunlight. Where the temperature changes extremely.
The equipment must be operated only by, or under direct supervision of a properly trained and qualified person. Customer maintenance of this equipment may only be performed as stated in the User’s Manual. Any additional maintenance may only be performed by HUVITZ service technicians or other authorized persons.
Page 11
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- The manufacturer is only responsible for effects on safety, reliability, and performance of this equipment when the following requirements are fulfilled: (1) The electrical installation in the respective room corresponds to the specifications stated in this manual and (2) This equipment is used, operated, and maintained according to this manual.
Aniseikonia tests. 34 Masks (vertical line, horizontal line, and single character) 2.5 ~ 8 m (8 ~ 26 feet) projection range Cordless Remote Control User Programmable (2 programs up to 30 charts each) Easily networked with HUVITZ’s Digital Refractor...
Don’t use organic solution such as alcohol, thinner, benzene, etc. to clean the surface of this instrument. It may damage the instrument. When moving the CCP-3100, always make sure the power supply is off, and then lift the bottom of the unit with both hands.
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual 5.Configurations 5.1 Main Body The CCP-3100 Chart Projector is mainly composed of the body, stand, screen and remote control. [Fig. 1 Front View with Remote Control] 1. Projection lens 2. IR receiving window 3. Stand (table mount)
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 5.2 Main Body cont. [Fig. 2 Bottom View] [Fig. 3 Rear View] 1. Focusing Wheel 2. Power ON/OFF LED 3.110/220 Volt Switch 4. Power On/Off Switch 5. Tilting Adjustment 6. RS-232 Comm.Port 7. CAN (Controller Area Network) 8. AC Power Inlet 9.
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 6.Installation 6.1 Position of Chart Projector The single tube projector must be installed at the same distance from the screen as the patient. [Fig. 6 Installation without mirrors] Install chart projector at desired position within 2.5~8m (8~26 feet).
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual 6.2 How to obtain correct Letter Size and focus (Zooming Type only) Install chart projector at desired position of 2.5~8m (8~26 feet) Measure the distance from patient to screen Refer to the test scale shown in the Appendix A Have someone hold the test scale at the screen Project the 20/400(6/120, 0.05) alphabet “C”...
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 7.Input and Use of Program 7.1 How to input user's program Step 1. Press the “LIGHT” button once. Results: The light goes out. Step 2. Press the “PGM 1” button three times. Results: The light will come on and the chart projector will display the first chart of the default program.
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual 7.2 How to use your stored programs. Step 1. Press “ PGM 1 ” or “PGM 2” to select the program of your choice. Step 2. Use the “Program ” or “Program ” to scroll through your stored...
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 8.Types of Chart (Type C is Standard) [Fig. 7 Types of Chart]...
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual 9.Maintenance and Repair Maintenance : There is no periodic or routine user maintenance required. 9.1 Lamp Replacement Turn off power and disconnect the power cord. Allow the projector to cool down for about 5 minutes before replacing the lamp.
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- [Fig. 8 Lamp assembly] Install the lamp securely as showen in the Fig.8 Make sure that you install it in the correct direction. 9.2 Fuse Replacement Turn off power, and disconnect the power cord. Pull out the plastic cover at the lower part of power socket, to replace the fuse.
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual [Fig. 9 Fuse assembly] 9.3 Replacement the Battery of Remote Controller Remove the battery cover of the remote controller. Replace the batteries and the battery cover to the remote controller. NOTE : Be sure to observe the polarity marks (plus + & minus - ) inside the battery compartment.
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 9.4 Cleaning There are no cleaning requirements other than regular office housekeeping, such as dusting. 1. Wipe the cover and the screen with a dry soft cloth when they get dirty. 2. If for some reason the projector should become soiled,wipe it clean with the damp lint-free cloth and mild detergent.
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- 11.Components List 11.1 Standard accessories 1. Remote control -1 2. Screen (400X500mm) -1 3. Table stand -1 4. Dust cover -1 5. Power cable –1 For use of equipment in rated voltage less than 125Vac,minimum 6A,Type SJT or SVT , 18/3AWG,10A, max 3.0m long : One end with Hospital Grade...
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual 12.Service Information How to contact service: If there are any problems with the equipment, please follow the steps below: First of all, refer to the 9. Maintenance and Repair sections according to the problem that you are encountered. And then follow the suggested sequences.
Page 29
CCP-3100 Operator’s Manual ----------------------------------------------------------------------- If you can’t contact with your local distributor, you can directly get in touch with the service department of the HUVITZ using the phone number and the address written in the below table. How to Contact HUVITZ Co., Ltd...
Page 30
-------------------------------------------------------------------- CCP-3100 Opreator’s Manual Appendix A Test Scale Hold this test scale at the Screen, and according to the refracting distance from the patient’s eye to the screen such as 3m, 4m, 5m or 6m position the projector backward and forward to fit the projector Alphabet 400 chart “C” with this Scale...
Need help?
Do you have a question about the CCP-3100 and is the answer not in the manual?
Questions and answers