Page 2
(dose, dose rate, low battery, defect, dose/dose rate overflow). The RAD-60 is ready for use after the battery has been installed. It must be worn in the chest area such as a chest pocket (pocket clip on the outside) with no metal objects in the same pocket.
Page 3
Switch off (OFF) Press push button shortly until -Display changes to OFF. Press push button until a bleep occurs. -The display goes blank. Chirp on/off (CHr) Press push button shortly until -Display changes to CHr. Press push button until a bleep occurs. -The chirp is activated/deactivated.
Page 4
Show dose rate alarm level (drA) Press push button shortly until -Display changes to drA. Press push button until a bleep occurs. -Current alarm level is displayed. Wait for return to dose/dose rate display. Change dose rate alarm level (drA) There are 6 different levels for dose rate alarm: 10 µSv/h, 100 µSv/h, 1 mSv/h, 10 mSv/h,...
Page 5
Dose alarm or Dose rate alarm (Blinking display) The dose/dose rate display is blinking and the continuos audible alarm is on. The dose alarm can be reset by pressing the push button. The display changes to previously selected mode. Audible alarms Dose rate: - - - - (4 bleeps each second) Dose: __...
Page 6
±15% Operational temperature range: -20°C to +50°C Operational humidity: 10 - 90% RH Dimensions: 78 * 67 * 22 mm, excluding pocket clip. Weight: 80g, including battery LAURUS Systems, Inc. - Ph: 410-465-5558 - Fax: 410-465-5257 - www.LaurusSystems.com...