Quickie Prelude Service Manual
©2007 Sunrise Medical Inc.
103748 Rev A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Quickie prelude and is the answer not in the manual?

Questions and answers

Summary of Contents for Sunrise Quickie prelude

  • Page 1 Quickie Prelude Service Manual ©2007 Sunrise Medical Inc. 103748 Rev A...
  • Page 2: Table Of Contents

    Quickie Prelude Service Manual Contents Introduction ............ 0.1 Section 4 VR2 Controller ..........0.2 Understanding VR2 Controller Diagnostics Plugs/Connectors .......... 0.3 Codes ............... 7 Basic Tool List & Main Wiring Diagram ..0.4 4.1 One Bar - Low Battery Voltage ) ....7 Section 1 4.2 Two Bars –...
  • Page 3: Introduction

    Service Manual. This Service Manual is intended as a troubleshooting guide for the Quickie Prelude. Pho- tographs and content may differ from the actual products in some cases due to changes in specifications and other factors.
  • Page 4: Vr2 Controller

    Speed/Profile indicator- A se- ries of five LED’s, which display speed and profile settings Speed/ Profile Decrease. Used Speed/ Profile Increase. Used to decrease the Speed/ Profile to Increase the Speed/ Profile setting. setting. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 0.2...
  • Page 5: Plugs/Connectors

    Motor Plug VR2 Controller M1 = RIGHT SIDE MOTOR M2 = LEFT SIDE MOTOR JSM = JOYSTICK MODULE INH-2 = INHIBIT 2 A1 = ACTUATOR 1 A2 =ACTUATOR 2 + - =BATTERY SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 0.3...
  • Page 6: Basic Tool List & Main Wiring Diagram

    • 19mm socket wrench • 5mm socket wrench Wiring Diagram for VR2 Controller Package A/C INPUT CHARGER 24v BRAKE 24v BRAKE SOLENOID SOLENOID RIGHT MOTOR LEFT MOTOR FUSE FUSE RIGHT LEFT CIRCUIT BATTERY BATTERY BREAKER SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 0.4...
  • Page 7: Troubleshooting: No Power

    Charge batteries until the voltage in each battery measures approximately 12 volts. Connect batteries back into chair and complete charge with standard on-board or off board charger. If the voltage meter reads below 12 volts, charge the batteries. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 1...
  • Page 8: Battery Connection Test

    A1.5.2 imately 24 volts, replace the controller. If it reads 12 volts or less proceed to the next step. A1.5.3 If all of the measurements read full voltage, then replace the controller. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 2...
  • Page 9: Circuit Breaker Test

    12 volts and polarity is correct, replace the main battery harness plugging into the controller. If voltage is less then 10 volts proceed to A1.6.1 section 1.3. If polarity is reversed correct battery wiring. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 3...
  • Page 10: Main Harness

    Next check the continuity between the white rear connector (figure 1.9.3) and the right contact on the controller plug. If the meter reads more than 1 ohm in any of the above tests replace the main harness. A1.9.3 SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 4...
  • Page 11: Chair Is Not Charging

    VR2 joystick is less then 18 volts (refer to section 1.3.) If the voltage is greater then 18 volts refer to you charger troubleshooting manual. A2.1.1 If all of the measurements read full voltage, then replace the controller. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 5...
  • Page 12: Understanding Vr2 Controller Display

    This indicates a system trip, i.e. the VR2 has detected a problem somewhere in the wheelchair's electrical system. Please refer to Section 3 (VR2 Controller Diagnostics). SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006...
  • Page 13: Understanding Vr2 Controller Diagnostics Codes

    A4.2.2 If none of the above corrects the problem, replace the left motor. If the meter reads between 0 to 1.5 ohms, then replace the controller. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 7...
  • Page 14: Three Bars - Right Motor Wiring Trip / Low Battery Voltage

    Test 3 Test 4 If all of the readings are open, then replace the controller. If any of the readings are short, then replace the left motor. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 8...
  • Page 15: Four Bars- Left Motor Disconnected

    A4.4.2 If the meter reads between 0 and 1.5 ohms, then replace the controller. If this does not correct the problem, then replace the right motor. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 9...
  • Page 16: Five Bars - Left Motor Wiring Trip

    Controller Fault - A control system trip is indicated. Make sure that all connections are secure. Check that the batteries are fully charged and in good condition, and check all joystick connections and cables. If this does not correct the prob- lem, then replace the controller. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006...
  • Page 17: Nine Bars - Solenoid Brake Trip

    (figure A4.10.1) If the meter reads more than 30 volts, then check the charger. Otherwise, replace your controller. If the meter reads more than 30 volts replace the charger, If the Batteries, connections, and voltage level are correct replace the controller. SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 11...
  • Page 18: Disassembly/Reassembly, And Adjustment

    Note: Remember to lay cables on the seat otherwise they may become tangled when taking off the seat. Unlatch the seat and rotate slightly (figure s1.2). Lift the seat verticaly to remove (figure s1.3). s1.1 s1.2 SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 12...
  • Page 19: Step 2 - Footrest Adjustment And Removal

    Remove the two attachment bolt ( fi g s2.1) Slide footrest assembly up or down to desired position and reinsert attachment bolts. s2.1 To adjust footplate angle turn screw in or out ( fi g s2.3) s2.3 SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 13...
  • Page 20: Step 3 - Shroud Removal

    Remove fender shroud . s3.3 Rear shroud Loosen and remove the two philips screws holding the rear shroud in place.(fi g. s3.4). Remove rear shroud . s3.4 SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 14...
  • Page 21: Step 4 - Battery Removal

    Disconect battery harnesses (fi g s4.1) . Remove front battery. Gently remove controller from velcro inside battery tray( fi g s5.1). Unplug all connectors from the controller. Remove controller from battery tray. s5.1 SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 15...
  • Page 22: Step 6 - Motor Removal

    Remove the cotter pin (A) (fig s6.4) holding the cog release handle (B) to the cog release lever arm (C). s6.1 Warning-If motor plate is removed form the gear box the gear box plate may come off. s6.2 s6.4 s6.3 SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 16...
  • Page 23: Step 7 - Caster Fork Removal

    Use two 13mm socket wrenches to remove the three wheel hub fasteners (fig s8.2) Separate the wheel hubs and remove the tire. Reassembly: Ensure the Motor Shaft Key is properly inserted and re- verse above instructions s8.1 s8.2 SUNRISE MEDICAL PRELUDE SERVICE MANUAL 2006 PAGE 17...