drop and test strip symbols. The WaveSense Presto System is
now ready for you to apply control solution. You do not need to
set a calibration code for this meter.
[3]
Applying Control Solution: Shake
the control solution bottle. Discard the
first drop of control solution and wipe
the bottle tip clean using a clean tissue
or paper towel. Dispense a second drop
onto a clean surface, such as an unused
resealable plastic bag. Bring the tip of the
test strip to the control solution sample.
The WaveSense Presto System will automatically detect if a
test is performed with WaveSense Control Solution.
Apply control
solution here.
When control solution is applied, the visual fill window on the
WaveSense Presto Test Strip turns blue.
18
This is the visual
fi ll window. DO
NOT PUT CONTROL
SOLUTION HERE.
The entire visual fi ll
window will turn blue.
Your meter will beep once indicating that you should remove
the tip of the WaveSense Presto Test Strip from the control
solution sample.
On the WaveSense Presto Meter display,
the droplet symbols appear on the test
strip symbol to indicate that the control
solution is filling the WaveSense Presto Test Strip.
[4]
Viewing Control Solution Test Result: The result will
appear on the display and, if the volume is enabled, your
meter will beep once. The word "control" will also appear on
the display.
Compare the result of your control
solution test to the range printed
on the WaveSense Presto Test Strip
vial label for the control solution
level (normal or high) that you
used. Your control solution result
should fall within this range. The
control solution test result is stored
in memory.
To deactivate the meter, remove
the used WaveSense Presto Test
Strip. Dispose of used materials
properly. If your control solution
test results are out of the range,
repeat the test. If the results from
the control solution tests continue
to fall outside that range, do not use
the WaveSense Presto System to test
your blood glucose. Call customer service.
Normal: 80-120 mg/dL
High: 220-320 mg/dL
19