Performing a Control Solution Test

The WaveSense Presto system has 2 levels of control solution:
normal and high. Select the level of control solution you want
to use.
The control solution is used to:
[1]
Ensure that your WaveSense Presto Meter and Test Strips
are working together properly.
[2]
Practice testing without having to use your own blood.
Control solution tests should be performed when you:
[1]
First get your WaveSense Presto Meter.
[2]
Begin using a new vial of WaveSense Presto Test Strips.
[3]
Suspect that your WaveSense Presto Meter or Test Strips
are not working properly.
[4]
Think your test results are not accurate.
[5]
Have dropped or damaged your meter or exposed your
meter to liquids other than blood samples.
[6]
Are advised by your healthcare professional to do so.
16
Control solution should only be used for 90 days after first
opening the bottle or until the expiration date printed on the
label, whichever comes first.
IMPORTANT: Count forward 90 days from the date you
open a new bottle of WaveSense Control Solution. This is
your discard date. Write the discard date on the WaveSense
Control Solution bottle.
Important WaveSense Control Solution information:
[1]
Use only WaveSense Control Solution with your
WaveSense Presto System.
[2]
Replace the cap on the control solution bottle
immediately after use.
[3]
Do not add water or any other liquid to control solution.
[4]
Control solution tests should be performed within the
system operating temperature range of 50 °F to 104 °F (10 °C
to 40 °C).
How to perform a control solution test:
[1]
Inserting a WaveSense Presto Test Strip: Gently insert
a new WaveSense Presto Test
Strip into the meter. The meter
will be activated.
[2]
Meter Ready For Test: The
meter now displays the blood
17