Application Information 3.1 General Information On Blood Pressure 4 Maintenance And Care 3.2 Purpose Of Use 7.1 Care 7.2 Calibration Of The „Klock“ Use Of The Device 4.1 Start Of Operation Warranty And Repair Conditions 4.2 How To Apply 8.1 Warranty The „Klock“...
1 Overview ! Start/Stop-button, described in text as left button @ Menu-button, described in text as right botton # Infrared-eye to transmit the stored measured data to the doctor’s computer $ Cuff % Battery compartment on the back ^ Display...
Germany greatest reliability. We thank you for your trust and are always available to answer questions I.E.M. GmbH declares that the product “Klock” about service and the product. corresponds to the guidelines according to 93/42/ECC of June, regarding Medical Products.
14 cm (5.5“) to 19.5 cm consciously adjust to a healthier lifestyle. (7.7“). The device is only suitable for blood • Please take your “Klock” with you on each pressure-measurements on adults. visit to the doctor. If your doctor is equipped •...
4 Use Of The Device 4.1 Start Of Operation 4.2 How To Apply The “Klock” To Your Wrist Open the battery compartment on the back The blood pressure measurement device is side of the device. Insert the batteries as shown put on like a normal wrist watch.
The above is only available if measurements have The measurement number will blink (here No.: been stored in the “Klock”. If not the display will 27). After 10 seconds this go directly to setting the time after pressing the measurement will be stored under this number right button (see: Setting Time And Date).
4.8 Reminder Function “Klock” is able to remind you, for instance, to take your medication. An audible signal will be heard. The setting of the different times (up to 5) can only be done at your doctor’s...
4 Use Of The Device 4.9 Description of Symbols Operation / Situation Meaning Pulse detected Display systolic value upper value SYS [mm ⁄ Hg] Display diastolic value lower value DIA [mm ⁄ Hg] Display pulse number of pulses / minute PULSE [ ⁄...
Please contact your local always consult with your physician to deter- service point. mine if the “Klock” Blood Pressure Watch is • The device should not be operated within safe for you to use. the area of a Nuclear Spintomograph or other highly sensitive electronic equipment.
6 Descriptions Of Errors And How To Correct Them 6.1 Error Messages Error number Cause Correction Err 01 Motion detected Keep arm, hand and fingers still and at the or Arrhythmia level of the heart repeat measurement Err 02 Cuff a) put cuff on correctly (see 4.2) b) the circumference of the wrist should be between 14 cm (5.5“) and 19.5 cm (7.7“)
6 Descriptions Of Errors And How To Correct Them 6.2 Trouble Shooting Description of error Correction of error The measurement cannot be stopped and the Open cuff, remove batteries, stop using cuff does not deflate instantly when pushing the device. Contact Service! the left button once Left button has been pushed but the measure- Check batteries, change if necessary...
7 Maintenance And Care 7.1 Care 7.2 Calibration Of The “Klock” If necessary, please clean the plastic housing The device has been calibrated by the and the cuff with a moist cotton cloth. manufacturer for a period of 2 years. We...
8 Warranty And Repair Conditions 8.1 Warranty In no event will I.E.M GmbH be responsible for See Warranty Card attached consequential damages resulting from the use I.E.M GmbH, the Manufacturer, warrants this of this product. product to be free of defects in material and workmanship for a period of one (1 year) from I.E.M.
Need help?
Do you have a question about the klock and is the answer not in the manual?
Questions and answers