5-D Power Connection and Operation Test

(1) Connect the power code after turning the power switch off.
(2) Turning the power switch on turns the "POWER" LED (orange) on to sound a buzzer.
This turns the device in the ready status for commands from the host computer.
If the "ERROR" LED (red) comes on or a buzzer continuously sounds, turn the power switch off
and refer to "5-F Troubleshooting".
(3) Pressing the "Feed" switch once feeds one card.
Keep pressing it continuously feeds cards.

5-E Test Mode

Test mode: Checks reading. For checking, use a specified check card.
(1) Turning the power switch on while pressing the "Feed" switch sounds a buzzer and calls the test
mode, turning all LEDs on.
(2) Pressing the "Feed" switch once continuously feeds cards. During this continuous feeding, a
buzzer goes off if a reading error occurs.
(3) Keep pressing the "Feed" switch stops card feeding.
(4) In order to exit from the test mode, turn the power off and on again.
Note
The check card to be used must be changed as follows depending on the setting of DIP
SW6:
DIP SW6 = OFF : Timing direct-under-type, 12-line, standard-sized card
ON : Timing control-type, 15-line, postcard-sized card
(With this setting, the mark-to-mark-type or direct-under-type selected by DIP
SW5 is neglected.)
-19-