Specifications 
LCD 
Size 
Elements 
Alarms 
Set points 
Reset options 
Selectable Contact Relay 
Contact rating 
Contact resistance 
Range 
Communications 
Method 
Speed 
Time accuracy 
Power 
Dimensions 
Weight 
Internal 12‐Bit Temperature Sensor 
Measurement range 
Accuracy 
Resolution 
Response time 
Temperature Accuracy Drift 
Internal 12‐Bit Relative Humidity Sensor (User replaceable) 
Measurement range 
Accuracy 
Resolution 
Response time 
Accuracy Drift 
Hysteresis 
External Sensors 
Battery Life 
 
Temperature Accuracy and Resolution 
The maximum reading error for the internal temperature sensors is shown at right. This is a worst case 
error. In a typical logger, temperature errors will be substantially lower. NIST certification of 
temperature accuracy for some sensors is available from Onset. The temperature resolution for internal 
sensors is also shown at right. Resolution is the smallest difference between adjacent temperature steps 
that the logger can report. The logger will typically experience little or no accuracy drift unless the 
electronic components or temperature sensor are damaged mechanically or by exposure to 
environmental factors beyond the specifications, such as high heat and humidity. 
HOBO U14 Data Logger User Manual
 
33 x 50.8 mm (1.3 x 2 inches) 
 Temperature and relative humidity simultaneously 
 °C or °F (selectable within host software) 
 Memory remaining and battery level 
 Flashing ALERT for out‐of‐limits conditions (selectable within host software), which stops flashing 
on offload or return to normal range 
High and low set points for both temperature and relative humidity with selectable delays 
Alert conditions reset on offload or return to normal range 
Normally Closed or Normally Open 
48VDC, 1A max 
Less than 1 Ohm 
0% Low Battery Range 
Offloads data to PC or U‐Shuttle via USB cable 
Readout full logger (64K) in less than 30 seconds 
±1 minute per month at +20°C (+68°F) 
3 AAA Alkaline batteries, user‐replaceable 
125 x 92 x 31 mm (4.9 x 3.6 x 1.2 in) 
170 g (6.0 oz) with batteries 
‐20°C to +50°C (‐4°F to +122°F) 
±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F); ±0.13°C (0.24°F) typical (see Plot A) 
0.02°C @ 25°C (0.04°F @ 77°F) (see Plot A) 
15 minutes (to 90% in airflow of 1 m/sec) 
Negligible 
0‐100% RH, ‐20 to 50°C (‐4 to 122°F) 
+/‐ 2.5% from 10% to 90% RH (typical), to a maximum of +/‐ 3.5% including hysteresis (see Plot B) 
0.03% RH @ 25°C (77°F) 
2 minutes (to 90% in airflow of 1 m/sec) 
<2% over 5 years typical 
1% Typical 
For External Sensor Specifications, see the sensor's user manual 
The batteries will last one year in typical conditions (logging intervals of ≥ 1 minute with weekly 
offloads and average temperatures greater than 10°C or +50°F). Frequent offloads and/or extreme 
temperatures will reduce battery life. At temperatures below freezing (0°C or +32°F), batteries will 
typically require replacement every six months. In these conditions the battery level bar graph 
should be checked regularly. See Battery Level Indicators on page 8. 
The CE Marking identifies this product as complying with all relevant directives in the European 
Union (EU). 
Plot A: Internal Temperature Sensor 