HOBO
U14 Data Logger User Manual
®
The U family of data loggers offers reliability and convenient monitoring for applications that 
require higher accuracy, better resolution, more memory, or USB connectivity for fast data 
readout to your computer. 
The HOBO U14 logger records and displays temperature and humidity conditions in 
manufacturing, processing, and storage environments where reliable monitoring and 
documentation of specific temp/RH conditions are critical. 
The HOBO U14 is available in two versions: U14‐001 with internal sensors and the U14‐002 
with external sensors. The HOBO U14 requires HOBOware
or later. 
Features 
 64k non‐volatile memory that retains data even if the batteries fail 
 Memory indicator on LCD shows how much data storage space remains 
 Battery Level indicator shows remaining useful battery life 
 Operates for one year on three alkaline AAA batteries in typical conditions 
 Built‐in relay for use with Remote Alarm or Automatic Dialer 
 Direct USB interface for launching and data readout using HOBOware 
 Sample (logging) interval can be set from one second to 18 hours 
 Supports Onset's Data Assistants (available with HOBOware software) 
 Can be used with the Onset U‐Shuttle (U‐DT‐1) 
 Internal sensor model (U14‐001) has internal temperature and relative humidity sensors with 12‐bit resolution 
 External sensor model (U14‐002) supports the 12‐bit Temperature/10‐bit RH Sensor (S‐THB‐M00x) and the 12‐bit 
Temperature Sensor (S‐TMB‐M00x) 
Accessories 
 Outdoor Case (CASE‐4x) 
 Remote Alarm (ARA) 
 Automatic Dialer (AVD‐45) 
Location Requirements 
The HOBO U14 is designed for a non‐condensing, non‐corrosive environment. Keep the logger dry. Some external sensors can be 
placed in harsher environments than the logger. See the external sensor specifications in the sensor user manual. 
Specifications 
Environmental 
Operating temperature range 
Operating relative humidity range 
Logger Capabilities 
Capacity 
Software‐selectable sampling intervals 
Start 
Recording modes 
Launch Modes 
Memory 
© 2010–2012 Onset Computer Corporation. All rights reserved. Onset, HOBO, and HOBOware are trademarks or registered trademarks of Onset Computer Corporation for its data logger 
products and configuration/interface software. All other trademarks are the property of their respective companies. 
13416‐C MAN‐U14‐00x 
®
 Pro or HOBOware Lite, Version 2.5 
 
‐20°C to +50°C, (‐4°F to +122°F) , 0 to 100% RH non‐condensing 
0 to 95%, non‐condensing, non‐corrosive environment 
Temp and RH: 21,500 measurements; Temp Only: 43,000 measurements 
1 second up to 18 hours, recording times up to 1 year 
Programmable start time/date 
Stop when full 
Immediate, Delayed 
Nonvolatile EEPROM memory retains data even if batteries fail 
U14 Data Logger  
(Model U14‐001 shown) 

Advertisement

loading

Summary of Contents for Hobo U14

  • Page 1 HOBO U14 Data Logger User Manual ® The U family of data loggers offers reliability and convenient monitoring for applications that  require higher accuracy, better resolution, more memory, or USB connectivity for fast data  readout to your computer.  The HOBO U14 logger records and displays temperature and humidity conditions in  manufacturing, processing, and storage environments where reliable monitoring and  documentation of specific temp/RH conditions are critical.  The HOBO U14 is available in two versions: U14‐001 with internal sensors and the U14‐002  ® with external sensors. The HOBO U14 requires HOBOware  Pro or HOBOware Lite, Version 2.5  or later.  Features   64k non‐volatile memory that retains data even if the batteries fail   Memory indicator on LCD shows how much data storage space remains  U14 Data Logger   (Model U14‐001 shown)   Battery Level indicator shows remaining useful battery life   Operates for one year on three alkaline AAA batteries in typical conditions   Built‐in relay for use with Remote Alarm or Automatic Dialer   Direct USB interface for launching and data readout using HOBOware   Sample (logging) interval can be set from one second to 18 hours   Supports Onset’s Data Assistants (available with HOBOware software)   Can be used with the Onset U‐Shuttle (U‐DT‐1)   Internal sensor model (U14‐001) has internal temperature and relative humidity sensors with 12‐bit resolution ...
  • Page 2: Specifications

    HOBO U14 Data Logger User Manual Specifications    LCD  Size  33 x 50.8 mm (1.3 x 2 inches)   Temperature and relative humidity simultaneously  Elements   °C or °F (selectable within host software)   Memory remaining and battery level   Flashing ALERT for out‐of‐limits conditions (selectable within host software), which stops flashing  on offload or return to normal range  Alarms  Set points  High and low set points for both temperature and relative humidity with selectable delays  Reset options  Alert conditions reset on offload or return to normal range  Selectable Contact Relay  Normally Closed or Normally Open  Contact rating  48VDC, 1A max  Contact resistance  Less than 1 Ohm  Range  0% Low Battery Range  Communications  Method  Offloads data to PC or U‐Shuttle via USB cable  Speed  Readout full logger (64K) in less than 30 seconds  Time accuracy ...
  • Page 3: Circuit Board

    HOBO U14 Data Logger User Manual Internal Relative Humidity Accuracy    Temperature compensated.    Do not expose the internal RH sensor to fog, mist, or other condensing  conditions.    The logger’s relative humidity accuracy for the internal model is shown in the  plot at right.   Plot B: Internal RH Sensor Accuracy Components  Front  Back  Mounting Holes. Use  LCD Screen.   mounting screws 2.5 cm  See The LCD  (1 inch) apart to mount  External Sensor  Screen section  the U14 to a wall. Port (Model  for details. U14‐002 Only).  Connect  USB Port.   supported  Use an Onset‐ Smart Sensor  supplied USB ...
  • Page 4 HOBO U14 Data Logger User Manual Setting Up the HOBO U14  Follow these instructions for setting up and launching the HOBO U14 logger.  1. Insert the three AAA batteries. Remove the battery door on the back of the unit and insert the batteries in the proper  direction.  2. If you are using a Remote Alarm and/or Automatic Dialer, connect it to the HOBO U14 now. Otherwise, replace the  battery cover and go to step 3.  The HOBO U14 is compatible with Onset’s Remote Alarm (part number ARA) and Automatic Dialer (part number AVD‐ 45) available from Onset Computer or an Onset Authorized Dealer. See the product’s user manual for complete details.  The interconnect wire range is 22AWG to 14AWG. To connect a Remote Alarm or Automatic Dialer to the HOBO U14:  Loosen the two screws on the relay  terminal. (Remove the battery door if  necessary.)  b. Feed the wires through the small hole at  the bottom of the case.  Feed the wires through the holes in the  relay terminal.  d. Tighten the screws on the terminal.  e. Replace the battery compartment cover.  Connect the wires to the Remote Alarm or Automatic Dialer using the following wiring diagrams as guides.  All three drawings show the logger in the Normally Closed state with the Alert inactive. The HOBO U14’s contact  closure is initially set at the factory to be Normally Closed, but you can change this in HOBOware to accommodate  Normally Open devices (see the ARA User’s Manual).  If the alert output is connected to a remote alarm configured for Normally Closed inputs, you can avoid false alarms  when the logger is disconnected by making a temporary connection between the wires going to the remote alarm.  Be sure to disconnect this temporary connection when the logger is reconnected to the remote alarm wires.    HOBO U14 with Onset’s AVD‐45 Automatic Dialer  HOBO U14 with Onset’s ARA Remote Alarm    ...
  • Page 5 HOBO U14 Data Logger User Manual HOBO U14 with Onset’s ARA Remote Alarm and AVD‐45 Automatic Dialer     NOTE: The polarity must be as shown. 3. If you have an External Sensor model (U14‐002): Plug the external sensor into the RJ‐12 jack on the side of the HOBO  U14. Hold the black rubber flap down against the connector so it slides into the port. NOTE: The sensor readings will not  be displayed until the logger is launched (immediate or delayed) or the status is checked using either HOBOware or the  U‐Shuttle. The readings will not be logged until you launch the logger later in this procedure.  4. Connect the U14 to a computer running HOBOware. Plug the large end of the USB cable supplied with HOBOware into   a USB port on the computer and plug the small end of the cable into the side of the HOBO U14. If the logger has never  been connected to the computer before, it may take a few seconds for the new hardware to be detected (Windows only).  5. Start HOBOware and open the Launch window. From the Device menu, select Launch. Configure Launch parameters,  including a Launch description, logging interval, and start time. Click Help in HOBOware for details on Launch  parameters.  6. Configure Alarms (optional). Use the Alarm feature to flash a warning on the LCD if the monitored temperature or  relative humidity falls outside user‐selectable limits. If the temperature or RH measured by the logger falls outside the  selected high or low limit, the word ‘Alert,’ an alert symbol, and the temperature or RH measurement that went out of  limits will flash.  7. If you require remote or secondary notification of an alarm condition, you can connect a remote audio alarm and/or an  Automatic Dialer to the contact closure as described in Step 2.  To access the Alarm feature, click the Alarms button in the Launch window, which opens the following dialog box.    5 ...
  • Page 6 HOBO U14 Data Logger User Manual Enable Sensor Alarms. To enable a sensor alarm, select the desired sensor in the Channel box, and then select the  check box for the High Alarm and/or the Low Alarm. Enter a value to define the alarm threshold. You can enter a  value in the box or use the sliders. (If you enter values manually, the software may adjust them slightly to the  nearest values supported by the logger.) You can also set the number of samples (30 seconds per sample) that must  be displayed before the alarm will be triggered.   b. Set Number of Samples. To avoid unintended activation of the relay contact, set the number of samples to allow  for anticipated breaches of the alarm threshold. If the alarm is tripped and the number of samples is set to more  than 1, the LCD will indicate Alert Delayed. If the measurement is still out of range when the number of samples is  reached, the relay contact will be activated.  Note about Avoiding Unintended Relay Activation. In some applications, the environment in which the logger  is launched is in immediate violation of the desired alarm limit and will cause an unintended opening or closing  of the contact relay. For example, setting an upper alarm limit of +32°F for a freezer application and then  launching the logger in a room at room temperature will cause an immediate alarm trigger. In this situation,  enable a Delayed Launch or select a number of samples that will result in a long enough delay to allow time to  transport, mount, and stabilize the logger in the desired location.  Other causes of invalid alarms would be freezer doors being opened frequently or defrost cycles. To avoid  these, set the number of samples to allow for small periods of time where the value is outside the range, but  where the alarm will not be tripped unless the value is outside the range for a significant number of samples.  Low Battery Alarm. Select the Battery Alarm checkbox to enable the Battery Alarm, which will trigger if the battery  drops to 0%.  d. Alarm Reset. Select whether the alarm will be deactivated when the Host has relaunched the logger or when the  sensor reading has returned within limits.  e. Set Relay Contact. If you are using an external device that will be triggered by an alarm (such as a Remote Alarm or  Automatic Dialer), set the desired default relay contact state to either Normally Closed (relay opens during an alarm  condition, which is the default) or Normally Open (relay closes during an alarm condition) The ARA Remote Alarm  can be set to Normally Closed or Normally Open (see remote device User’s Manuals for more details). NOTE: The  AVD‐45 Automatic Dialer requires the logger’s contacts to be set to Normally Closed.  Test the Relay Alarm. Use the Test Set Alarm button to test the relay alarm function. When you click the Test Set  Alarm button, the logger will activate the alarm and change the condition of the contact closure. When you click  the Test Set Alarm button, it changes to the Test Clear Alarm button. Click the Test Clear Alarm button to  deactivate the alarm. ...
  • Page 7: The Lcd Screen

    HOBO U14 Data Logger User Manual In some applications (such as a refrigerator, freezer, clean room) it may be more convenient to permanently mount an  external sensor at the site. Be sure to place the remote sensor in a location that is representative of the target  measurement location and allows free flow of air (beware of any drafts caused by vents, doors, etc).   If you permanently mount the sensor and later have to disconnect it from the logger for readout, select Delayed Launch  when you relaunch the logger to allow time to reconnect the sensor before the actual launch time.  13. Read out the HOBO U14 as described in the next section. When the memory is full, or anytime you want to view the  data, read out the HOBO U14 using HOBOware.  Reading Out the HOBO U14  NOTE: If you are using the U‐Shuttle to read out the logger, see the U‐Shuttle User Manual for specific instructions.   To offload data from the HOBO U14:  1. Optional: Disconnect the logger from the external sensor, Remote Alarm or Automatic Dialer.  2. Transport the logger to the location of the computer.  3. Connect the logger to the computer.  4. Open HOBOware software on the computer.  5. Optional: If you have more than one logger connected to the computer, connect to the HOBO U14 by selecting Device >  Select and then select the HOBO U14 from the Select Device window.  6. From the main menu, click the Readout icon (or select Device > Readout from the main menu).  7. If the logger is currently logging, you can choose to continue logging while offloading or stop logging. The readout will  begin. The status bar in the Readout Logger window will indicate the percentage complete.   8. When prompted, enter a name for the file and click Save. The Plot Setup Dialog appears.  9. Click Plot to view the graph of the data. See the HOBOware User Manual or On‐line Help for more details on graphing,  time selection in plots, and exporting data files.  10. To redeploy the logger, re‐launch it with a delayed start allowing enough time to return the logger to the site, reconnect ...
  • Page 8 HOBO U14 Data Logger User Manual  Temperature Reading. This displays the current temperature (Fahrenheit or Celsius as selected in HOBOware) on the  LCD. Internal sensor has temperature measurement range of ‐20°C to +50°C (‐4°F to +122°F). Notes: There may be slight  variations between the displayed temperature value and the logged value due to limits in display conversions. The  values seen in HOBOware and the U‐Shuttle are the actual recorded values and the most accurate. The displayed  temperature on the LCD screen may be ±0.3°C (±0.6°F) different than the logged value. Fahrenheit values of 200° and  higher are displayed as three digits without a decimal (e.g., 205°F); however, the full resolution value will be recorded in  the logger memory (e.g., 205.4).   Relative Humidity (RH) Reading. This displays the current RH on the LCD. The internal sensor has a measurement range  of 0% to 100%.   Alarm Icon. This displays alerts as configured within HOBOware. Specifically, ALERT indicator signals when temperature  or relative humidity limits have been exceeded. ALERT DELAYED if you have delay enabled and an alarm is tripped.  Triangle with exclamation mark immediately after delay (if enabled) indicates that the relay is in an alarm state.   Battery Level. Remaining battery level is indicated by a five‐segment bar. The number of segments decreases as battery  level decreases. Batteries should be replaced when the BATT LEVEL bar graph is down to one segment, which indicates  20% battery life or less.   Remaining Memory. Remaining memory is indicated by a five‐segment bar. The number of segments decreases as  remaining memory decreases. NOTE: If the memory is full, the logger will no longer log. The logger will continue to  monitor conditions however, and will trigger an alert if the reading falls outside of the configured range. You must  readout the logger and then relaunch it using HOBOware to resume logging.  LCD Indicators If the temperature or RH measured by the logger falls outside the selected high or low  Alarms limit, the word ‘Alert,’ an alert symbol, and the temperature or RH measurement that  went out of limits will flash. If you set the number of samples to greater than 1 and the  measured value is outside the limits, ALERT DELAYED will flash until either the number  of samples is reached (ALERT with a triangle symbol is displayed) or the alert condition  no longer exists. Battery Level BATT LEVEL bar graph down to one segment indicates 20% battery life or less. Batteries ...
  • Page 9 HOBO U14 Data Logger User Manual LCD Indicators Memory Display States  State  MEM LEFT Hash  Segments  Illustration  marks  Stopped  On  None    Launched/  On  Number of segments depends  Logging  on memory remaining. Highest  segment flashes.    Launch  Flashing  All On/Flashing  Delaying    Memory Full  Flashing   All Off  (until data is  downloaded)        External Sensor Removed If an external sensor is disconnected from the logger, the logger will record values of ‐...
  • Page 10: Maintenance

    HOBO U14 Data Logger User Manual Maintenance  Replacing Batteries  The HOBO U14 requires three AAA batteries. Onset Computer Corporation strongly recommends the use of fresh alkaline  batteries. Batteries should be replaced when the BATT LEVEL bar graph is down to one segment, which indicates 20% battery life  or less. To ensure data integrity, use HOBOware to read out the logger and stop logging before changing batteries.   Replacing the Internal RH Sensor (Model U14‐001 Only)  The internal relative humidity sensor in the U14‐001 logger may require replacement if it is damaged mechanically or by  exposure to condensation or chemicals. The replacement sensor (Part No. HUM‐RHPCB‐3A) is available from Onset or an Onset  Authorized Dealer for U14‐001 loggers with serial numbers above 9847173. Note: U14‐001 loggers with a serial number below  9847173 require a different replacement sensor (Part No. HUM‐RHPCB‐3). You can replace the internal RH sensor yourself or  have it replaced as part of a U14‐001 tune‐up/calibration service (available from Onset). Contact Onset for more details.                                               ...

This manual is also suitable for:

U14 seriesU14-002U14-001