Results screen

Linearity Test
12:48 pm
Linearity 12345678 L1
Strip Lot 545794
09/18/13
The hourglass icon indicates the test is running. When the
test is completed and the result is ready, the meter beeps
again.
Linearity Test
12:48 pm
Linearity 12345678 L1
Date 09/18/13 12:48 pm
45
mg/dL
09/18/13
You can add comments to a test result (as with blood
glucose tests, see page 52).
If you do not wish to add a comment to the test result,
touch
to continue to the next level of the linearity
test.
Remove the test strip and dispose of it in accordance with
applicable regulations and directives. Repeat the above
steps for all levels of the linearity test.
Test results are also saved when the meter is powered
off in the results screen or when it automatically
powers off.
Linearity Testing
Linearity Test
12:48 pm
Linearity 12345678
Strip Lot 545794
L1
L2
L3
L4
L5
L6
09/18/13