Aeroneb Go Instruction Manual

Aeroneb Go Instruction Manual

Micropump nebulizer
Hide thumbs Also See for Go:

Advertisement

Instruction Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Go and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Aeroneb Go

  • Page 1 Instruction Manual...
  • Page 3: Table Of Contents

    Table of Contents Introduction............1 Warnings............2 Warranty............6 Classifications.and.Markings......7 Aeroneb.Go.Parts..........8 Assembly/Functional.Test........9 Operational.Instructions........10 Cleaning.and.storing........12 Troubleshooting.Guide........15 Electromagnetic.Susceptibility......16 Specifications..........17 Performance............ 18 Replacement.Parts.and.Accessories....19...
  • Page 4 This page has been intentionally left blank...
  • Page 5: Introduction

    The.Aeroneb . Go,. for. use. by. paediatric. and. adult. patients,. ® is. intended. to. aerosolize. physician-prescribed. solutions. for. inhalation. that. are. approved. for. use. with. a. general-purpose.
  • Page 6: Warnings

    The.AC/DC.adapter.does.not.charge.batteries. •. If.you.are.using.batteries.and.the.green.light.is.flashing,. your.batteries.are.low..You.may.finish.your.treatment.and. then. change. (or. replace. and. recharge). the. three. “AA”. batteries.immediately. •. Do. not. share. your.Aeroneb . Go. Nebulizer. with. others.. ® It.is.intended.for.use.by.a.single.user..If.more.than.one. person. uses. it,. there. is. a. risk. of. spreading. infectious. illness. •. As.with.all.electronic.devices,.it.is.recommended.that.it. the.product.is.unplugged.when.not.in.use.
  • Page 7 •. If.after.proper.cleaning.it.takes.longer.than.12.minutes.to. deliver.3mL.of.normal.saline.(0.9%),.the.Nebulizer.Unit. should.be.replaced. •. The. user. should. ensure. that. exhaled. medication. is. directed.away.from.bystanders. •. Do.not.microwave.any.parts. •. By. connecting. the.AC/DC. adapter. to. the.Aeroneb . Go. ® Control. Module,. the. Aeroneb Go. Control. Module. and. ®. batteries.are.not.being.charged. •. As.with.any.cables.keep.all.cables.tidy.to.avoid.tripping. or.strangulation.hazards.and.take.particular.care.around. children.
  • Page 8 •. Do.not.use.brushes,.sharp.tools.or.abrasive.cloths.that. could.scratch.or.damage.the.surfaces. •. Do.not.clean.parts.in.a.dishwasher. •. Store.your.Aeroneb .Go.in.the.carrying.case. ® •. Do. not. store. the.Aeroneb . Go. in. a. location. where. it. is. ® exposed.to.direct.sunlight,.extreme.heat.or.cold,.dust.or. water..Store.out.of.reach.of.children. •. Follow.local.laws.and.recycling.plans.regarding.disposal. or.recycling.of.components,.batteries.and.packaging. •. Before.storing.your.Aeroneb .Go.for.an.extended.period,. ® remove. the. batteries. from. the. Aeroneb Go. Control.
  • Page 9: Warranty

    WARRANTY The. Aeroneb . Go. Control. Module. and. AC/DC. Adapter. are. ® warranted.for.a.period.of.two.years.from.the.date.of.purchase. against.defects.in.manufacturing..The.Aeroneb .Go.nebulizer. ® unit.is.warranted.for.one.year.from.date.of.purchase.against. defects.in.manufacturing..All.warranties.are.based.on.typical. usage. The.life.of.the.Aeroneb .Go.nebulizer.unit.has.been.validated. ® for. use. for. 100. disinfecting. treatments. (boiling. or. baby. bottle. steam. steriliser).. This. is. based. on. a. minimum. of. one.
  • Page 10: Classifications.and.markings

    CLASSIFICATIONS AND MARKINGS The.symbols.that.appear.on.the.Aeroneb .Go.Nebulizer.and. ® its.accessories.are.described.in.the.table.below: Symbol Meaning Attention,.consult.accompanying.documents Class.II.equipment.per.IEC.60601-1 Type.BF.equipment.per.IEC.60601-1..This.label. applies.to.Aeroneb .Go.nebulizer.unit.only ® Classified.by.TUV.with.respect.to.electric.shock,.fire. and.mechanical.hazards Serial.number Model.number Manufacturer Input Output DC.Voltage AC.Voltage Power.On/Off Fragile.–.Handle.with.care Keep.Dry Storage.temperature.limits:.-20°C.to.60°C –20 This.device.compiles.with.the.requirements.of.the. Medical.Devices.Directive.(93/42/EEC) Consult.Instructions.for.Use...
  • Page 11: Aeroneb.go.parts

    Control.Module AC/DC.Adapter* Mask.Adapters *. A vailable.with.complete. Aeroneb .Go.System ® UNIT ASSEMBLY AND USAGE NOTE:. Clean. all. parts. of. your. Aeroneb . Go. Nebulizer. ® (described.on.page.11).before.first.use,.after.each.use,.or.after. extended.storage. PRIOR TO ASSEMBLY: Check. all. parts. for. visible. damage.. Replace. any. damaged. parts.before.you.use.the.Aeroneb .Go..In.the.case.of.missing.
  • Page 12: Assembly/Functional.test

    ASSEMBLY/FUNCTIONAL TEST 1.. Connect. the. Medication. Cup. to. the. Nebulizer.Body.by.lining.up.the.notches. on. the. medication. cup. and. nebulizer. body. NOTE: Make sure the Aeroneb Go is ® unplugged before you add medication to the Medication Cup. 2.. Insert.the.nebulizer.cable.into.the.back.of. the.nebulizer.unit. 3.. Add.Aeroneb .Go.base.to.the.bottom.of.
  • Page 13: Operational.instructions

    (up. to. five. (5). hours. from. high. performance. batteries). 2.. Connect. the. nebulizer. cable. to. the. Aeroneb . Go. Control. Module,. indicated. ® by.the.Output.symbol.( 3.. Turn.on.the.unit.by.pressing.the.“On/Off”. button. on. the. Control. Module.. You. will. see.a.green.light.on.the.Control.Module.
  • Page 14 5.. Alternatively,. you. can. power. the. device. by. using. the.AC/DC.Adapter. if. provided. (refer. page. 7).. Plug. the.AC/DC. adaptor. into. a. standard. AC. electrical. outlet,. a. green.light.on.the.adapter.indicates.that. the.AC/DC.adapter.is.turned.on. 6.. Plug. the. AC/DC. cable. into. the. Control. Module,. indicated. by. the. DC. Voltage. Symbol.( NOTE: the AC/DC adapter does not charge batteries.
  • Page 15: Cleaning.and.storing

    CLEANING AND STORING YOUR AERONEB ® AFTER EVERY USE: Disconnect.the.control.module.before.you.take.the.nebulizer. apart.to.rinse.it. 1.. Rinse. the. Nebulizer. Unit. (Medication. Cup,. Nebulizer. Body,. Base. and. Mouthpiece. or. Mask. Adapter). under. running.hot.tap.water.after.every.use. 2.. Shake.off.excess.water,.and.allow.parts.to.fully.air.dry.on. a.clean,.dry.towel,.out.of.the.reach.of.children. 3.. Make. certain. that. all. cleaned. parts. are. completely. dry.
  • Page 16 OPTIONAL DISINFECTION: If. you. would. like. to. disinfect. your. Aeroneb . Go. Nebulizer,. ® you. may. use. choose. one. of. two. options;. Boiling. or. Vinegar. Solution. BOILING: Rinse. the. Nebulizer. Unit. (Medication. Cup,. Nebulizer. Body,. Base. and. Mouthpiece. or. Mask. Adapter). under. running. hot. tap. water.
  • Page 17 OPTIONAL DISINFECTION (cont’d): VINEGAR SOLUTION: Rinse. the. Nebulizer. Unit. (Medication. Cup,. Nebulizer. Body,. Base. and. Mouthpiece. or. Mask. Adapter). under. running. hot. tap. water. after.every.use. ii.. Shake.off.excess.water,.and.allow.parts.to.fully. air.dry.on.a.clean,.dry.towel,.out.of.the.reach. of.children. iii.. Fill.a.container.with.a.mixture.of.three.(3).parts. hot.water.to.one.(1).part.white.distilled.vinegar. iv.. Soak.the.Nebulizer.Unit.parts.(Medication.Cup. and. Cap,. Nebulizer. Body,. and. Mouthpiece. or.
  • Page 18: Troubleshooting.guide

    TROUBLESHOOTING GUIDE If.you.have.difficulty.operating.your.Aeroneb .Go,.check.the. ® following.points: Problem Check Remedy No visible Check.that.the. Using.the.markings. aerosol batteries.are. under.each.battery. when using inserted.correctly. location,.place. batteries each.battery.in.the. Control.Module. housing..Match.(+). and.(-). Check.to.see.if.the. Replace.the. low.battery.power. batteries.or.use.the. light.is.flashing. AC/DC.Adapter. Check.that.the. Connect.the.cable. Control.Module. properly.to.the. cable.is.connected. nebulizer.and.the. properly.to.the. Control.Module. nebulizer.and.to.the.
  • Page 19: Electromagnetic.susceptibility

    Medication.Cup. medication. source is Check.that.the. Clean.the. functioning nebulizer.has.been. Aeroneb .Go..Refer. ® properly.cleaned. to.page.11. If.above.remedies.do.not.work,.call.your.local.supplier. ELECTROMAGNETIC SUSCEPTIBILITY This. device. meets. the. requirements. of. EN60601-1-2. which. describes. requirements. for. electromagnetic. compatibility. (EMC). of. medical. devices.. This. includes. immunity. to. radio.
  • Page 20: Specifications

    SPECIFICATIONS POWER.SOURCES: Aeroneb .Go.Control.Module:.4.5.VDC.(3.“AA”.disposable.or. ® rechargeable.batteries),.350.mA AC/DC.Adapter:.100-240.VAC,.200.mA,.50-60.Hz POWER.CONSUMPTION: Less.than.2.watts APPROXIMATE.WEIGHT: Aeroneb .Go.Nebulizer.Unit:.60g ® Aeroneb .Go.Control.Module.with.batteries:.260g ® AC.Powered.Controller:.180g AC/DC.Adapter:.100g SOUND.LEVEL: <35.dBA.at.1.meter APPROXIMATE.DIMENSIONS: Aeroneb .Go.Nebulizer.Unit.(assembled): ® 40.0.x.105.0.x.95.0.mm Aeroneb .Go.Control.Module: ® 70.0.x.110.0.x.32.0.mm AC/DC.Adapter: 74.(H).x.72.(D).x.49.(L).mm RECOMMENDED.OPERATING.ENVIRONMENT: Temperature.range:.5°C.to.+45°C.(+41°F.to.+113°F) Atmospheric.pressure:.450.to.1100.mBars Humidity:.up.to.95%.relative.humidity RECOMMENDED.TRANSPORT.AND.STORAGE. ENVIRONMENT Temperature.range:.-20°C.to.+60°C.(-4°F.to.+140°F) Atmospheric.pressure:.450.to.1100.mBars Humidity:.up.to.95%.relative.humidity...
  • Page 21: Performance

    PERFORMANCE: Aerosol.flow.rate:.Greater.than.0.3.mL.per.minute Average.tested.flow.rate:.0.4.mL.per.minute Residual.Volume:.<0.1.mL.for.a.3.ml.dose Aerosol.output.rate:.0.25.mg.per.minute.(salbutamol) Aerosol.Output:.0.86.mL Average.particle.size:.3.5.microns.mass.median. aerodynamic.diameter Medication.Cup.capacity:.Maximum.of.6.mL The.temperature.of.the.medication.will.not.rise.more.than. 10°C.(50°F).above.ambient.temperature.during.normal.use. Performance.may.vary.depending.on.the.type.of.drug.and. nebulizer.unit.used..For.additional.information.contact. Aerogen.or.your.drug.supplier. Representative.particle.size.distribution.graph.for.albuterol. as.per.EN13544-1.is.shown.below: 100.0 50.0 0.1. Aerodynamic Size (µm)
  • Page 22: Replacement.parts.and.accessories

    REPLACEMENT PARTS AND ACCESSORIES: PART REORDER NUMBER Aeroneb .Go.System AG-AG4200-XX* ® Medication.Cup.Cap AG-AG3010 Aeroneb Go.Handset AG-AG4300 ® Mouthpiece AG-AG3040 Handset.Base.(pack.of.10) AG-AG4030 AC/DC.Adapter* AG-AG3410-XX* Control.Module AG-AG3060 Control.Module.Cable AG-AG4070 Mask.Adapter.(22mm.M) AG-AG3080 Mask.Adapter.(22mm.F) AG-AG3090 Carrying.Case AG-AG3110 Instruction.Manual* AG-AG4120-XX* *. C ontact. your. local. supplier. for. reorder. number. specific. for.
  • Page 24 . Aerogen.Limited, Galway.Business.Park, Dangan, Galway, Ireland. www.aerogen.com Ph:.+353.91.540400 ©2010.Aerogen.Limited AG-AG4120-EN.Rev.B.2010. Manufacturing.no.:.30-553.Rev.B...

Table of Contents