Troubleshooting the Typhoon

6.1

Troubleshooting the Typhoon with ACS

If a problem occurs with the wheelchair, then please proceed as follows:
• Identify the possible cause of the fault by using the troubleshooting tables below.
• Check the Status Display on the remote. Identify the error code if it is flashing.
• Perform the necessary checks and repairs as recommended in the table below.
Note
For more information on troubleshooting the Typhoon with GB motors, please see the "Dynamic
DX-GB-AS Power Module - Installation Manual", Order Nr. 1441533
6.1.1

Diagnosing driving faults

PROBLEM
Wheelchai
r will not
drive
OTHER
POSSIBLE
SYMPTOMS
CAUSE
Status display
Drive motors
on remote
may be
lights up
disengaged
normally and
does not show
an error code
Status display
Batteries may
on remote
be defective
does not light
up
Batteries may
be completely
discharged
Power supply
to the remote
may be
interrupted
Remote may
be defective
Status display
Various causes
on remote is
flashing
SOLUTION
REFERENCE
See User Manual
Engage
the drive
motors
See "Replacing the
Replace
batteries" on page 33
the batteries
See User Manual
Charge
the batteries
See "Replacing the main
Check
fuse" on page 39
the main fuse
See "Checking the
Check
cables" on page 41
cables
between
modules for
loose
connections or
damage
See "Replacing the ACS
Exchang
Remote" on page 42
e the remote
on the
wheelchair for
a different one
to eliminate
the possibility
that the
remote may
be the cause.
See "REM24 Error
Identify
Codes and Diagnostic
the error code
Codes" on page 16