Summary of Contents for Thermo Scientific Alpha pH 600

  • Page 1 Instruction Manual Alpha pH 600 RF Transmitter System pH / ORP / ºC / ºF 68X450201 | Rev.0 | Oct 2008...
  • Page 3 Preface This manual serves to explain the use of the Alpha pH 600 Series RF transmitter. It functions in two ways, firstly as a step by step guide to help you to operate the transmitter. Secondly, it serves as a handy reference guide. It is written to cover as many anticipated applications of the transmitter as possible.
  • Page 5: Table Of Contents

    Electrode Offset Settings ................24 Temperature Settings..................25 Buffer Selection Settings................27 Configuration Settings..................28 Viewing Electrode Properties................. 31 TECHNICAL SPECIFICATIONS LIST OF ACCESSORIES Thermo Scientific ................... 33 Eutech Instruments ..................34 TROUBLESHOOTING GENERAL INFORMATION Warranty......................36 Return of Goods..................... 36 APPENDICES 10.1...
  • Page 7: Introduction

    They will be glad to assist you. Intended Use The Alpha pH 600 Series RF Transmitters are intended solely for pH or ORP and temperature measurements, as described in this instruction manual. This transmitter system is currently designed for use in 923 to 928 MHz license-free ISM radio band transmission, which is licensed free in the following countries only.:...
  • Page 8: Safety Instructions

    Remove the unit from the wall. Correct Disposal of the Instrument When the Alpha pH 600 Transmitter is permanently taken out of service, obey the local environmental regulations for correct disposal or send the instrument to your local distributor, they will take care of proper disposal.
  • Page 9: Getting Started

    GETTING STARTED Description of Instrument The Alpha pH 600 RF Transmitter is used for measuring pH or ORP and temperature values. The pH values can be measured using industrial combination pH sensors; the ORP values can be measured using industrial ORP sensor; the temperature values can be measured using 3-wire Pt100 / Pt1000 sensors.
  • Page 10: Measurement System

    Measurement System A typical measurement system consists of: Up to 10 Alpha pH 600 Transmitters Alpha Link 600 Data Acquisition Software (Installation instructions included separately) A Wireless RF Transceiver One pH sensor or ORP sensor with integrated or separate Pt100/Pt1000...
  • Page 11: Connecting Peripherals

    Connecting Peripherals 2.3.1 Connection Terminals Remove Back Cover: • Remove the screws from the four corners at the back of the pH Transmitter. • Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below. Connectors: •...
  • Page 12 2.3.2 Switching Between PT100 & PT1000 Temperature Sensors The Alpha pH 600 transmitter supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J201) as described below.
  • Page 13 2.3.3 Connecting pH/ORP Electrode If the pH/ORP electrode has a BNC connector, remove the BNC connector from the cable. NOTE: Oakton Instruments offers an optional ‘BNC to Spade Lug adapter’ (Order code: 05994-90) that can be used with pH/ORP electrode without removing the BNC connector. Strip the insulation of the cable so that the bare wires are exposed enough for connection as shown in Figure 3.
  • Page 14 2.3.4 Connecting Temperature Probe For Automatic Temperature Compensated (ATC) pH readings, a 100Ω Pt RTD temperature probe (2-wire or 3-wire) can be connected to the Controller. 3-Wire Probe: Connect PT100 compensate wire to Pin 5 of J204 connector Connect PT100 sense wire to Pin 6 of J204 connector Connect PT100 GND wire to Pin 7 of J204 connector 2-Wire Probe: Short Pin 5 &...
  • Page 15: Installation

    Installation 2.4.1 Mechanical Dimensions Alpha pH 600 PH/ORP RF Transmitter 2.4.2 Wall Mount Pierce through holes at both sides Cover the catch slots at both sides with overlays - 9 -...
  • Page 16: Display & Keypad

    Display & Keypad 2.5.1 Display Overview The Liquid Crystal Display (LCD) of Alpha pH 600 Transmitter has two alpha- numerical displays (Upper and a Lower). • Upper display: Measured pH, mV or relative mV value are displayed when Transmitter normal operation (measurement) mode.
  • Page 17 Mode Indicators Measurement mode; blinks in Symmetric mode MEAS (Refer Section 3.1& 5.5 ) SETUP Setup mode (Refer Section 5) Calibration mode (Refer Section 4) Status Annunciators READY Appears when the reading is stable Appears when Automatic Temperature Compensation (ATC) is enabled.
  • Page 18 2.5.2 Key Functions Alpha pH 600 PH/ORP RF Transmitter Description Brings directly into the Calibration entry mode. The security code must be selected to enter the calibration mode or view only mode. While in pH Measurement mode, press CAL to enter pH Calibration entry mode.
  • Page 19: Operation

    OPERATION Measurement mode When the pH Transmitter is powered on, the display shows all the LCD segments briefly, and then automatically enters into the Measurement mode. MEAS READY 7 .00 °C 2 5 . 0 The mode indicator ‘MEAS’ at the top of the display indicates that the pH Transmitter is in Measurement mode;...
  • Page 20: Menu Overview

    Menu Overview 3.2.1 Alpha pH 600 Calibrat See Sectio MEAS °C SETUP Offset Set tings See Sect ion 5.2 SETUP Temperatur e Settings See Sect ion 5.3 SETUP View Electrode Properties See Secti on 5.6 SETUP Buffer Selectio n Settings See Sectio n 5.4...
  • Page 21: Calibration Mode

    CALIBRATION MODE Preparing the Transmitter & Electrode for Calibration Before starting calibration, make sure that the pH Transmitter is in appropriate measurement mode (pH or ORP). When the Transmitter is powered on, it starts up with the measurement mode last used. For example, if the pH Transmitter is powered off in ORP measurement mode, it starts up in ORP mode when it is powered on.
  • Page 22: Ph Calibration

    READY °C • To exit Calibration mode at any time press ▲ and ▼ keys simultaneously (Escape). • The calibration is always carried out in the units of HOLD measurement (pH or mV), selected in setup mode pH Calibration The pH Transmitter is capable of calibration of up to 2 points using USA or NIST pH buffer standards.
  • Page 23: Point Calibration

    From pH measurement mode press CAL key to enter calibration mode as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to begin calibration. Place the electrode in pH 7.00 buffer. Immerse the temperature probe in the buffer solution if ATC mode is enabled. Immerse the potential matching pin in the buffer if symmetrical mode is enabled.
  • Page 24 when the reading is stable. Press ENT key to confirm the reading. The pH Transmitter moves to the second calibration point. The lower display shows next standard buffer value (pH 4.01). Use ▲ and ▼ keys to select your second buffer from one of the preset values: pH 4.01 or 10.01.
  • Page 25 1-Point Calibration: MEAS READY READY HOLD HOLD HOLD °C °C From pH measurement mode press CAL key to enter calibration mode as described in section 4.2. The LCD shows ‘CAL PH’. Press ENT key to begin calibration. Place the electrode in pH 6.86 buffer. Immerse the temperature probe in the buffer solution if ATC mode is enabled.
  • Page 26 pin in the buffer if symmetrical mode is enabled. The buffer annunciator appears in LCD. Lower display shows pH 6.86 (NIST standard buffer). Upper display shows the current uncalibrated pH reading. Allow the reading to stabilize. LCD shows ‘READY’ annunciator when the reading is stable.
  • Page 27: Check Calibration

    Check Calibration MEAS READY While in Measurement mode press the CAL key. The display will prompt you to enter a security code. Leave the security code at “0” (view only mode). Press the ENT key. The display shows the slope in mV (Main Reading) and the normalized temperature is shown on the Auxiliary reading of the display.
  • Page 28: Temperature Calibration

    Press the ENT key to confirm. The transmitter calculates the correction factor and returns to the Measurement mode. Temperature Calibration Calibrate temperature probe only if temperature value displayed on the pH Transmitter is different from that of a calibrated thermometer. Refer to section 5.3 for further information temperature settings.
  • Page 29: Setup Mode

    S E T U P M O D E Enter Setup mode In the Setup mode the transmitter can be configured to the individual requirements. MEAS “User-setting” Mode READY (User can change parameter settings) °C “User-Setting” Mode (User can only view parameters) While in measurement mode, press the ENT key.
  • Page 30: Electrode Offset Settings

    Electrode Offset Settings NOTE: ‘Electrode Offset Setting’ is not available when the pH Transmitter is configured for ORP measurement mode. (Refer Section 5.5 for switching measurement modes) In applications where continuous pH measurement is required, it may not be convenient to remove the electrode for calibration. In such cases, an on-line offset adjustment is recommended.
  • Page 31: Temperature Settings

    ‘READY’ annunciator when the reading is stable. Press ▲ or ▼ key to set the upper display to the newly calculated offset value. As the upper display value (offset) changes, the pH Transmitter adjust the lower reading automatically to suit the new offset value. Up to ± 2.00 pH offset is allowed.
  • Page 32 From pH or ORP measurement mode press ENT key to enter setup mode as described in section 5.1. The LCD shows the first screen of setup mode (OFS). Press ▲ or ▼ key to select Temperature settings screen (SET ºCF). Press ENT key to access temperature settings (SET ºCF).
  • Page 33: Buffer Selection Settings

    Setting calibration temperature: The lower display shows ‘C.ºC’ and the upper display shows the last configured calibration temperature. Press ▲ or ▼ key to adjust the upper display to desired calibration temperature. Allowable range: –10.0 to 110.0°C / 14.0 to 230°F. Press ENT key to confirm the calibration temperature.
  • Page 34: Configuration Settings

    From pH measurement mode press ENT key to enter setup mode as described in section 5.1. The LCD shows the first screen of setup mode (OFS). Press ▲ or ▼ key to select buffer settings screen (bUFF). Press ENT key to access buffer settings (bUFF). Selecting buffer group: The lower display shows the last configured buffer group.
  • Page 35 SETUP SETUP SETUP SETUP SETUP SETUP SETUP SETUP SETUP Measurement Mode SETUP HOLD NOTE: To exit from any intermediate steps, press ▲ and ▼ keys (Escape) together. It will return to measurement mode. Selecting “pH” or “mV” Mode of Operation Press ENT key and use ▲...
  • Page 36 In such cases, switch the Transmitter to the Symmetrical mode as below: Press ENT key and use ▲ or ▼ keys to scroll till LCD displays ‘CONF’ (Configuration). Press ENT again. LCD displays “ASY” (asymmetrical operation). Use ▲ or ▼ keys to select for ‘”SY”...
  • Page 37: Viewing Electrode Properties

    Viewing Electrode Properties Each time you calibrate your pH electrode, the pH Transmitter re-calculates slope & offset of the electrode and shown in the LCD at the end of calibration. The setup mode allows you to view slope & offset values at any time. NOTE: When the measurement mode is set to ORP, only the offset value of the electrode is displayed.
  • Page 38: Technical Specifications

    T E C H N I C AL S P E C I F I C AT I O N S General Specification (a) pH pH Measuring Range 0.00 to 14.00 pH Resolution 0.01 pH Accuracy ± 0.01 pH (b) mV mV Measuring Range -1000 to 1000 mV Resolution...
  • Page 39: List Of Accessories

    Thermo Scientific pH Transmitter Replacement and Accessories Item Description Order Code Alpha pH 600 wall-mount pH/ORP RF transmitter TSPHCTP0600 Wireless (RF) receiver and application software TSRCV0600S Ryton®-body pH combi electrode with Pt100 RTD (ATC) & 20 m cable with BNC &...
  • Page 40: Eutech Instruments

    Eutech Instruments pH Transmitter Replacement and Accessories Order Code Item Description Alpha pH 600 transmitter 56706-00 Wireless RF receiver with application software 56706-50 pH/Temp electrode with PMPO and 10-ft cable 35807-20 Platinium ORP electrode with 10-ft cable 35801-21 BNC to spade lug adapter...
  • Page 41: Troubleshooting

    T R O U B L E S H O O T I N G Cause Problem Solution Power on, but no display Loose connections Ensure cables make good contact Incorrect output voltage of the power adaptor Use an power adaptor with specified output voltage Unstable pH reading Dirty electrode...
  • Page 42: General Information

    If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight pre-paid – and correction will be made without charge. Thermo Scientific/ Eutech will determine if the product problem is due to deviations or customer misuse.
  • Page 43: Appendices

    APPENDICES 10.1 Appendix 1– pH Buffer Values at Various Temperatures The following table shows the various pH values at different temperature of the solution during calibration. pH 4.01 pH 6.86 pH 7.00 pH 9.18 pH 10.01 4.01 6.98 7.12 9.47 10.32 4.01 6.95...
  • Page 44: Appendix 2 - Abbreviations Used In Lcd

    10.2 Appendix 2 – Abbreviations Used in LCD Abbreviation Description Asymmetrical mode Automatic Temperature Compensation bUFF Buffer Calibration C.ºC Calibration temperature CONF Configuration Default values Factory (defaults) Offset Reading is over range Out of range Oxidation Reduction Potential P.ºC Process temperature Setting Symmetrical mode Type...
  • Page 48 Thermo Scientific Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 www.thermo.com/process Europe Denmark House, Angel Drove Ely, Cambridgeshire CB7 4ET, UK Tel: 44-1353-666111 Fax: 44-1353-666001...