  • Page 1 Alpha COND 500 Controller/Transmitter uS / mS / ºC...
  • Page 2 ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
  • Page 3 Preface This manual serves to explain the use of the Alpha COND 500 transmitter. It functions in two ways, firstly as a step by step guide to help you to operate the transmitter. Secondly, it serves as a handy reference guide. It is written to cover as many anticipated applications of the transmitter as possible.
  • Page 4: Table Of Contents

    TABLE OF CONTENTS INTRODUCTION ................... 1 PREPARATION ..................2 (SL2 P )........... 3 OWER UPPLY EQUIREMENTS OSITION (SL1 P ) ..3 ONNECTING THE LECTRODE AND EMPERATURE ENSOR OSITION INSTALLATION ..................5 ................5 ECHANICAL IMENSIONS ....................5 OUNT .................... 6 ANEL OUNT DISPLAY AND KEYPAD FUNCTIONS ..........
  • Page 5: Introduction

    INTRODUCTION Thank you for selecting an Alpha COND 500 2-Wire Transmitter. This isolated output 4-20mA transmitter is a sturdy microprocessor-based instrument that measures conductivity and temperature and transmits its output via the 2-wire power supply loop. This transmitter has many user-friendly features – all of which are completely accessible through the water-resistant membrane keypad.
  • Page 6: Preparation

    PREPARATION Remove screws from the four corners at the back of the Transmitter, and remove back cover. Connectors should be exposed as follows: Figure 1 – Connection Guide All wiring is done on 2 detachable connectors: – 9-pin connector (located on SL1 position) for Conductivity electrode and temperature sensor;...
  • Page 7: Power Supply Requirements (Sl2 Position)

    Power Supply Requirements (SL2 Position) The Alpha COND 500 transmitter requires a 12 to 24V DC power supply. Other Transmitters and/or a chart recorder may be connected in the loop. Insert positive loop wire from power supply to pin 1, tighten screw. Insert negative loop wire to pin 2, tighten screw.
  • Page 8 2.2.2 To connect ATC temperature sensor: For Automatic Temperature Compensated (ATC) Conductivity readings, an in-built temperature sensor is usually integrated in the industrial conductivity electrodes. Otherwise, a separate 100Ω Pt RTD temperature probe can be used. 3 wire temperature sensor Insert the Pt100 input wire to pin 5 of SL1 connector.
  • Page 9: Installation

    INSTALLATION Mechanical Dimensions Alpha COND 500 2-wire Conductivity Transmitter Wall Mount...
  • Page 10: Panel Mount

    Panel Mount 1. Prepare panel cut-out of Panel (side) 92.0 mm by 92.0 mm 2. Remove back cover of transmitter and slide it through panel cut-out 4. Thread rods through lugs until transmitter is held in place against panel 3. Atta ch lugs to either side of transmitter...
  • Page 11: Display And Keypad Functions

    DISPLAY AND KEYPAD FUNCTIONS Display The LCD has a primary and secondary display. • The primary display shows the measured conductivity value. • The secondary display shows the measured temperature. In Calibration mode, measured conductivity values are displayed here. Primar y Display (Upper Display)
  • Page 12: Keypad

    Keypad Alpha C OND 500 MEAS MEAS READY READY μS μS °C °C 2-wire Conductivity Transmitter 2-wire Conductivity Transmitter The four-button keypad allows easy and quick operations of the Transmitter. Function Brings you directly into the Calibration mode. If you were in Conductivity Measurement mode, press CAL to enter Conductivity Calibration mode.
  • Page 13: Calibration

    CALIBRATION Important Information on Transmitter Calibration Calibration should be carried when you are using your transmitter with a new electrode for the first time or when you suspect that the transmitter/electrode is out of calibration. Your transmitter allows you to perform temperature calibration and conductivity calibration.
  • Page 14 Table 1 Conductivity Recommended Calibration Solution Measuring Range Range 19.99 μS 6.00 to 17.00 μS 0.00 199.9 μS 60.0 to 170.0 μS 1999 μS 600 to 1700 μS 0.00 10.00 mS 6.00 to 7.00 mS 0.00 19.99 mS 6.00 to 17.00 mS 199.9 mS 60.0 to 170.0 mS To view current calibration point, see Section 6.8 - Viewing Calibration .
  • Page 15: Temperature Calibration

    Temperature Calibration You need to perform temperature calibration if your transmitter temperature reading is inaccurate or if the transmitter’s cell constant setting is changed. It is because the transmitter’s temperature offset calibration will be erased once a new cell constant is selected.
  • Page 16: Conductivity Calibration

    Conductivity Calibration MEAS A 1-point Calibration is required for this transmitter. If 1578 μS the calibration process is aborted, your transmitter will revert to the previous calibration data. °C Organize your calibration standard solution 25.5 in two beakers – one for rinsing and the other for calibration.
  • Page 17: Advanced Setup Functions

    ADVANCED SETUP FUNCTIONS The advanced setup mode lets you customize your transmitter’s preferences and defaults. This transmitter features different sub groups that organize all setup parameters. The sub-groups are: Range and Zooming Selection Setting Your COND 500 transmitter provides six selections of conductivity measurement ranges to suit your process application needs.
  • Page 18 For zooming setting, use the ▲ or ▼ keys to set SETUP 0.00 the conductivity low zoom value for the 4 mA μS HOLD current and press the ENT key to confirm. The transmitter now displays the high zoom value setup.
  • Page 19: Temperature Compensation Setting

    Temperature Compensation Setting Conductivity readings are affected by temperature. Under varying temperature conditions, use ATC to compensate for the conductivity values. If temperature of sample is constant, and a temperature sensor/probe is not available, Manual Temperature Compensation can be utilized. 6.2.1 Automatic Temperature Compensation For automatic temperature compensation (ATC) selection, connect the ATC sensor...
  • Page 20 6.2.2 Manual Temperature Compensation For manual temperature compensation you can set the process and calibration temperatures. This allows calibration at a temperature other than the process temperature. Example: setting a calibration temperature of 25°C lets you calibrate using standard solutions at 25°C, even if your process temperature is different from 25°C.
  • Page 21 SETUP HOLD SETUP HOLD SETUP HOLD °C Temperature Compensation Setting Chart – Alpha COND 500 Transmitter...
  • Page 22: Hold Current Setting

    HOLD Current Setting SETUP When Transmitter is in CAL or SETUP modes, it HOLD automatically goes into a ‘HOLD’ mode. To indicate Transmitter is in ‘HOLD’ mode, output current can be set to 22 mA output by activating the ‘HLD On’. Press the ENT key and use ▲...
  • Page 23: Temperature Coefficient And Normalization Temperature Setting

    Temperature Coefficient and Normalization Temperature Setting Since different process liquid may require different temperature coefficient factor for its temperature compensation calculation, COND 500 transmitter allows 0 to 10% temperature coefficient factor adjustment to cater for your different application needs. You also have the selection of 25.0 °C or 20.0°C for the conductivity measurement normalization temperature.
  • Page 24: Cell Constant Setting

    Cell Constant Setting You can select up to three types of cell constant in the setup mode: - K=1.0, K=0.1 or K=10. From the measurement mode, SETUP Press the ENT key to enter the setup mode. Use HOLD the ▲ or ▼ keys to scroll till LCD displays ‘SEt’ in the upper display;...
  • Page 25: Viewing Calibration Point

    Viewing Calibration Point This mode lets you view the current calibration point and its range. From the measurement mode of a selected measuring SETUP range, HOLD Press the ENT key to enter the setup mode. Use the ▲ or ▼ keys to scroll till LCD displays ‘CAL’ in the upper display.
  • Page 26: Reset Function

    6.10 Reset Function SETUP The reset function lets you choose to either reset the HOLD transmitter’s conductivity calibration only or conductivity calibration plus all setup functions that you might have changed back to factory default settings. However, the temperature compensation setting, as in Section 6.2, will remain unchanged.
  • Page 27: Factory Default Settings

    FACTORY DEFAULT SETTINGS SETUP DEFAULT FUNCTION PAGE SETTING Selection of effective conductivity range 1999 uS, r3 SET rng Setting low zooming selection 0 uS, Lo Setting high zooming selection 1999 uS, Hi Selection of Automatic or Manual SET °C AtC On Temperature Compensation Selection of HOLD current to be of 22 mA or SET HLd...
  • Page 28: Trouble Shooting Guide

    TROUBLE SHOOTING GUIDE Problem Cause Solution Power on but a). Loose connections a). Ensure cables are making no display good contact. b). Cables not in correct polarity (+ and – position). b). Re-wire loop cables with correct polarity. Unstable a). Air bubbles in probe. a).
  • Page 29: Specifications

    SPECIFICATIONS Cell Constant Conductivity Range Resolution (r1 to r6) 0.1, 1.0 Range 1 = 0 – 19.99 uS/cm 0.01 uS 0.1, 1.0, 10.0 Range 2 = 0 – 199.9 uS/cm 0.1uS 0.1, 1.0, 10.0 Range 3 = 0 – 1999 uS/cm 1 uS 0.1, 1.0, 10.0 Range 4 = 0 –...
  • Page 30: Accessories

    10 ACCESSORIES Replacement Unit Thermo Scientific Eutech Product Description Code No. Code Code No. Alpha COND 500 2-Wire Conductivity TSCONCTP0500 ECCONCTP0500 Transmitter Assembly Accessories Product Description Order Code Conductivity cell, Epoxy body, Graphite sensor, w/3-wire ECCONSEN89X Pt100, k = 0.1 Conductivity cell, Epoxy body, Graphite sensor, w/3-wire ECCONSEN88X Pt100, k = 1.0...
  • Page 31: Warranty

    11 WARRANTY This transmitter is supplied with a one-year warranty against significant deviations in material and workmanship from date of purchase and a six-month warranty for probe. Each instrument will have a warranty card with a specific serial number. The warranty card must be endorsed by the Authorized Distributor at the point of sale.
  • Page 32 Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832 Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949...