Easymax MU User Manual page 23

Easymax mu self-monitoring blood glucose system
Hide thumbs Also See for EasyMax MU:
Table of Contents

Advertisement

Troubleshooting Check
Was the test strip exposed to open air for a long
period of time?
Does test strip cap close tightly? Or was test
strip cap left open?
Is the meter functioning well?
Is the control solution expired or contaminated?
Were test strips and control solutions stored in
cool, dry places?
Did you follow the testing steps properly?
Action
If yes, repeat the control test with properly stored
strips.
If the cap was not tight, or the bottle was left
uncapped, open a new bottle of test strips.
Do not reuse the strips from the affected bottle.
You can use the control solutions to verify the
meter's functions.
If yes, replace with a new control solution to check
the performance of SMBG system.
If no, repeat the control test with properly stored
strips or control solutions.
Read Chapter 2 "Control Solution Testing" and test
again. Stop using the meter if you continue to
obtain the inaccurate results.
23

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EasyMax MU and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF