Omron compA-I-R NE-C28P Instruction Manual

Omron compA-I-R NE-C28P Instruction Manual

Compressor nebulizer
Hide thumbs Also See for compA-I-R NE-C28P:

Advertisement

Compressor Nebulizer
NE-C28P
Model
• Instruction Manual
• Mode d'emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
• РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ
TM
IM-NE-C28P-E-01-09/2011
EN
FR
DE
IT
ES
NL
RU
AR
5334069-0A

Advertisement

Table of Contents
loading

Summary of Contents for Omron compA-I-R NE-C28P

  • Page 1 Compressor Nebulizer NE-C28P Model • Instruction Manual • Mode d’emploi • Gebrauchsanweisung • Manuale di istruzioni • Manual de instrucciones • Gebruiksaanwijzing • РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ IM-NE-C28P-E-01-09/2011 5334069-0A...
  • Page 2: Table Of Contents

    Before using the unit Contents Before using the unit Introduction ....................3 Intended use ....................3 Important safety instructions ..............4 1. Know your unit ..................7 2. Guide to using the COMP AIR .............. 9 Operating instructions 3. Preparing the nebulizer for use ............10 4.
  • Page 3: Introduction

    • The user should also be capable of understanding general operation of NE-C28P and the content of instruction manual. Intended Patients This product should not be used by patients, who are unconscious or are not breathing spontaneously.
  • Page 4: Important Safety Instructions

    Important safety instructions Read all the information in the instruction manual and any other literature included in the box before using the device. When using an electrical product, especially when children are present, basic safety precautions should always be followed including the following: Warning: Indicates a potentially hazardous situation which, if not avoided, could result in serious injury.
  • Page 5 Important safety instructions • Do not immerse the compressor in water or other liquid. • Do not use or store the device in humid locations, such as a bathroom. • Do not operate the device with a damaged power cord or plug. •...
  • Page 6 • Plug the device into the appropriate voltage outlet. Do not overload power outlets or use extension cords. • Do not wind the power cord around the compressor. • Changes or modifications not approved by OMRON HEALTHCARE will void the user warranty. • Do not pull the power cord strongly.
  • Page 7: Know Your Unit

    1. Know your unit Rear View A. Compressor (Main Unit) D. Inhalation Attachments A1.Power Switch D1.Mouthpiece A2.Air Tube Connector D2.Adult Mask (PVC) A3.Air Filter Cover D3.Child Mask (PVC) (* Air filter inside.) A4.Nebulizer Kit Holder D4.Nosepiece A5.Ventilation Slots E. Nebulizer Kit A6.Power Cord E1.Cap A7.Power Plug...
  • Page 8 1. Know your unit Accessories Compressor Nebulizer NE-C28P Model • Instruction Manual • Mode d’emploi • Gebrauchsanweisung • Manuale di istruzioni • Manual de instrucciones • Gebruiksaanwijzing • IM-NE-C28P-E-01-09/2011 5334069-0A F. Spare Air Filters x 5 H. Instruction Manual G. Carrying Bag...
  • Page 9: Guide To Using The Comp Air

    2. Guide to using the COMP AIR Preparing the Nebulizer for Use - Section 3 (page 10-13) Warning: Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or masks before using them for the first time after purchase, if the device has not been used for a long period of time, or if more than one person uses the same device.
  • Page 10: Operating Instructions

    Operating instructions 3. Preparing the nebulizer for use Warning: Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or masks before using them for the first time after purchase, if the device has not been used for a long period of time, or if more than one person uses the same device.
  • Page 11 3. Preparing the nebulizer for use Add the correct amount of prescribed medication to the medication tank. Caution: The capacity of the medication tank is 2 to 7 ml. Note: Scales on the nebulizer kit are for reference only. The scale on the outside of the tank is with no vaporiser head in place.
  • Page 12 3. Preparing the nebulizer for use Attach the desired inhalation attachment, as described below. Attaching the mouthpiece Attach the cap to the inhalation air inlet. Attach the mouthpiece to the inhalation top. Attaching the adult mask (PVC) Attach the mask to the inhalation top. Note: If the amount of aerosol is too great, attach the cap.
  • Page 13 3. Preparing the nebulizer for use Attach the air tube. Twist the air tube plugs slightly and push it firmly into the air tube connectors. Notes: • Make sure that the air tube is securely attached to the compressor (main unit) and nebulizer kit, and does not come loose.
  • Page 14: Taking Medication

    4. Taking medication Hold the nebulizer kit as indicated on the right. Caution: Do not tilt the nebulizer kit at an angle of greater than 45° Right angle in all directions. Medication 45° 45° may flow into the mouth. Press the power switch to the on ( ) position.
  • Page 15 4. Taking medication Inhale medication according to your doctor’s and/or health care adviser’s instructions. Using the mouthpiece Insert the mouthpiece into your mouth and inhale the medication breathing normally. Exhale as normal through the mouthpiece. Using the nosepiece Inhale the medication through the nosepiece, as shown and exhale through the mouth.
  • Page 16 4. Taking medication When treatment is complete, turn the power off. Check that there is no condensation or moisture in the air tube and unplug the device from the power outlet. Power Plug Warning: Condensation may occur in the air tube. Do not store the air tube while it contains condensation or moisture.
  • Page 17: Care And Maintenance

    Care and maintenance 5. Cleaning and daily disinfecting Cleaning Following the cleaning instructions after each use will prevent any remaining medication in the tank from drying, resulting in the device not nebulizing effectively and also for preventing infections. Warning: Thoroughly rinse the nebulizer kit and the mask, nosepiece or mouthpiece with clean hot tap water after each use, and disinfect the parts of the nebulizer kit and applicable inhalation attachments after the last treatment of the day.
  • Page 18: Disinfecting The Nebulizing Parts

    5. Cleaning and daily disinfecting If the compressor is dirty, wipe it clean with a soft cloth moistened with water or mild detergent. Caution: The compressor is not waterproof. If the outside of the air tube is dirty, wipe it clean with a soft cloth moistened with water or mild detergent.
  • Page 19 5. Cleaning and daily disinfecting Handling the Vaporiser Head The vaporiser head is an important part that is used to vaporise medication. Be sure to observe the following precautions when handling it. Caution: • Always wash the vaporiser head after each use. Baffle •...
  • Page 20 5. Cleaning and daily disinfecting...
  • Page 21: Maintenance And Storage

    6. Maintenance and storage To keep your device in the best condition and protect the device from damage follow these directions: Caution: • Do not leave the unit unattended with infants or persons who cannot express their consent. • Do not subject the device or the parts to any strong shocks such as dropping the device on the floor.
  • Page 22: Changing The Air Filter

    Do not boil the air filter cover. Remove the air filter cover, wash it, and make sure it is dry and reattached correctly. • Please contact your OMRON dealer for additional air filters, or if the parts supplied with the device are damaged or missing.
  • Page 23: Troubleshooting

    The device is very hot. blocked? slots. Note: If the suggested remedy does not solve the problem, do not try to repair the device - no parts of the unit are user serviceable. Return the unit to an authorized OMRON retail outlet or distributor.
  • Page 24: Technical Data

    Notes: • Subject to technical modification without prior notice. • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan. • The device may not work if the temperature and voltage conditions are different to those defined in the specifications.
  • Page 25 9. Technical data Technical data for the OMRON NE-C28P Compressor with the OMRON V.V.T. (Virtual Valve Technology) Nebulizer kit N-01: Particle Size: *MMAD **approximately 3µm Medication Tank 7 ml maximum Capacity: Appropriate Medication 2 ml minimum - 7 ml maximum...
  • Page 26 Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter. Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual. Documentation is also available at www.omron-healthcare.com.
  • Page 27: Accessories

    10.Accessories Omron V.V.T. Nebulizer Kit Air Tube (PVC, 200cm) Mouthpiece N-01 Set Model: NEB-TP-L20E Model: C28-3P-E Order No.: 9520606-3 Order No.: 9956273-5 Model: C801S-NSET3-E Order No.: 9515573-6 Mouthpiece Air Tube (Silicone, 100cm) Nosepiece Model: C28-6-E Model: C28-7-E Order No.: 9956269-7 Order No.: 9956274-3...
  • Page 28 10. Accessories Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com OMRON (DALIAN) Co., Ltd. Production facility Dalian, CHINA OMRON HEALTHCARE UK LTD. Opal Drive Fox Milne, Milton Keynes, MK15 0DG, U.K.

Table of Contents