2. Guide to using the COMP AIR

Preparing the Nebulizer for Use - Section 3 (page 10-13)
Warning:
Clean and disinfect the nebulizer kit, mouthpiece and nosepiece or
masks before using them for the first time after purchase, if the
device has not been used for a long period of time, or if more than
one person uses the same device.
Taking Medication - Section 4 (page 14-16)
Caution:
• Do not tilt the nebulizer kit at an angle of greater
°
than 45
in all directions.
• Do not crease or fold the air tube.
Cleaning Parts After Use - Section 5
(page 17-20)
Note: Failure to clean the nebulizing parts
after use may lead to infection.
How the nebulizer kit works
The medication that is pumped up
through the medication channel is
mixed with compressed air which is
generated by a compressor pump.
The compressed air mixed with
medication is tuned into fine
particles and sprayed by being
impacted to the baffle.
Baffle
Aerosol
Medication
Nozzle
Channel
Medication
Medication
Compressed Air
Aerosol