2G Tarsys Systems

While the wheelchair is moving, minimum ground clearance for the front rigging is
three inches. If the wheelchair is not moving, the front rigging MUST maintain a
minimum of one inch ground clearance - otherwise personal injury and damage may
result.
NOTE: For this procedure, refer to FIGURE 6.17.
1. Remove the two mounting screws that secure the footrest extension tube to the
extension tube housing.
2. If adjusting to one of the three shortest settings for footrest height, remove the
retaining ring on the angle adjustment pin and slide it out.
3. Adjust the footrest extension tube to the desired height and align the corresponding
holes to the mounting holes on the extension tube housing.
4. If adjusting to one of the three shortest settings for footrest height, install the angle
adjustment pin and the retaining ring.
5. Reinstall the two mounting screws to secure the footrest extension tube to the
extension tube housing. Securely tighten.
6. Repeat STEPS 1-5 for the other extension tube.
Extension Tube Housing
Extension Tube
FIGURE 6.17 Adjusting the Height of the Center Mount Footrest - 2G Tarsys Systems
Part No 1114809
WARNING
Retaining Ring
SECTION 6—FRONT RIGGINGS
Seat Frame
Mounting Screws
Extension Tube Mounting Holes
Height Adjustment Holes
79
Release Lever
Angle Adjustment Pin